Comparison of the Parker flex tip and the unoflex reinforced endotracheal tube for orotracheal fibreoptic intubation in simulated difficult intubation patients

Rupesh Narhari
1, 2
,
Wan Mohd Nazaruddin Wan Hassan
1, 2
,
Rhendra Hardy Mohamad Zaini
1, 2
,
Sanihah Che Omar
1, 2
,
Nik Abdullah Nik Mohamad
1, 2
,
Praveena Seevaunnamtum
1, 2

1.
Department of Anaesthesiology and Intensive Care, School of Medical Sciences, Universiti Sains Malaysia (USM), Kubang Kerian, Kelantan, Malaysia
2.
Hospital USM, USM, Kubang Kerian, Kelantan, Malaysia
Anaesthesiol Intensive Ther 2020; 52, 5: 377–382
Online publish date: 2020/12/04
View full text
Get citation
ENW
EndNote
BIB
JabRef, Mendeley
RIS
Papers, Reference Manager, RefWorks, Zotero
AMA
APA
Chicago
Harvard
MLA
Vancouver
 
Introduction
The choice of endotracheal tube (ETT) is important for successful orotracheal fibreoptic intubation (OFI). The aim of this study was to compare the use of the Parker flex tip (PFT) with the unoflex reinforced (UFR) ETT during OFI.

Material and methods
A total of 58 patients who underwent elective surgery under general anaesthesia were randomised to two ETT groups, the PFT group (n = 29) and the UFR group (n = 29), for OFI in simulated difficult intubation patients using a rigid cervical collar. After successful standardised induction and relaxation, OFI and railroading of selected ETT were subsequently performed by a similarly experienced practitioner. Ease of insertion, degree of manipulation, time to successful intubation, post-intubation complications and haemodynamic changes were recorded for both groups.

Results
The percentage of easy intubation was comparable between both groups with a slightly higher percentage in the UFR group than the PFT group (69.0% vs. 62.0%; P = 0.599). Degree of manipulation was also comparable between the two groups; the percentage of cases in which manipulation was not required was slightly higher in the UFR group than the PFT group (69.0% vs. 62.1%; P = 0.849). Time to successful intubation was also comparable between the groups, although the time was slightly shorter for the UFR group than the PFT group (56.9 s ± 39.7 s vs. 63.9 s ± 36.9 s; P = 0.488). There were also no significant differences in other parameters.

Conclusions
The Parker flex tip ETT was comparable to the unoflex reinforced ETT for OFI in simulated difficult airway patients.

keywords:

fibreoptic, endotracheal tube, general anaesthesia, intubation