title="Special paperDouble stent loss during multivessel percutaneous coronary intervention in non-ST-segment elevation acute coronary syndrome",
abstract="The following description presents a case of a 75-year-old female patient with non-ST-segment elevation acute coronary syndrome in whom there occurred an infrequent complication of percutaneous coronary angioplasty: uncontrolled intracoronary stent displacement from a balloon catheter. During the attempt to retrieve the device using the “small-balloon” technique, further slippage of the stent from the balloon catheter to the right deep femoral artery was observed. Therefore, it was decided to provide a commercial intravascular microloop, whereby the stent was successfully retrieved outside the vascular system.",
author="Hawranek, Michał
and Desperak, Piotr D.
and Gąsior, Paweł M.
and Piegza, Jacek
and Ciślak, Aneta
and Lekston, Andrzej
and Gąsior, Mariusz",
pages="114--118",
doi="10.5114/pwki.2014.43518",
url="http://dx.doi.org/10.5114/pwki.2014.43518"
}