Postępy w Kardiologii Interwencyjnej

Abstract

1/2019 vol. 15
Original paper

Clinical experience with 12-month follow-up in patients after implantation of a novel long-tapered sirolimus drug-eluting stent

Adv Interv Cardiol 2019; 15, 1 (55): 46–51
Online publish date: 2019/03/15
View full text

Introduction

Long lesions contribute to a significant number of percutaneous coronary interventions

Aim

To assess the efficacy and safety of a novel long-tapered drug-eluting stent (DES) at a 12-month follow-up (FU) in patients with long coronary atherosclerotic lesions

Material and methods

A prospective clinical cohort study was conducted in 32 patients who underwent percutaneous coronary intervention using a BioMime Morph tapered stent (Meril Life Sciences, India). The patients were followed for 3, 6, and 12 months. The safety endpoints were death, myocardial infarction (MI), target lesion revascularization (TLR), target vessel revascularization (TVR), and MACE and/or major bleeding.

Results

Mean lesion length was 48 mm (range: 35–70 mm) measured via quantitative coronary analysis (QCA). In most cases, the target lesion was located in the LAD (68.75%). A GuideLiner catheter (Vascular Solutions Inc., MN, USA) was used in 12.5% of procedures; buddy-wire technique in 9.4% of cases. Bifurcation lesions were treated in 40.6% of cases. Additional stent implantation was needed in 56% of the procedures (25% of cases due to proximal or distal dissection, or due to insufficient stent length in 31% of cases). On 12-month FU we observed 1 TLR (3.1%), 1 TVR (3.1%), and 1 non-cardiovascular death.

Conclusions

The long sirolimus-eluting stent with tapered structure was characterized by good deliverability in long coronary lesions, although in some cases “buddy wire” or extension microcatheter use was necessary. Follow-up at 3, 6, and 12 months showed no significant major adverse cardiovascular events related to the device.

Share
without publication fees