Runaway pacemaker syndrome: a forgotten phenomenon or a forgotten pacemaker?

  1. Second Department of Arrhythmia, Cardinal Stefan Wyszynski National Institute of Cardiology, Warsaw, Poland

Kardiochirurgia i Torakochirurgia Polska 2026; 23 (3): 295-297

Data publikacji online: 2026/10/12
Article file
Runaway pacemaker.pdf

Runaway pacemaker syndrome (RPS) is a rare phenomenon in which a pacemaker (single or dual-chamber) induces potentially lethal ventricular arrhythmias as a result of device malfunction due to low battery voltage [1]. Pacemaker-induced ventricular tachycardia and loss of pacing bradycardia is a clinical entity that requires urgent attention. Battery depletion can result in low-rate stimuli, capture and sensing failures, mode changes, and, less frequently, RPS [2]. In RPS, abnormal device activity may produce rapid pacing that is not appropriately inhibited by the patient’s intrinsic rhythm. The resulting tachycardia may be refractory to conventional treatment if the underlying device malfunction persists [3]. The ECG may reveal a paced rhythm with ventricular pacing at a rate of 150–200/min, or captured beats alternating with non-captured high-rate spikes. This medical emergency should be considered in patients with old pacemakers, particularly when unexplained tachyarrhythmias or pacing abnormalities occur. Cardiothoracic surgeons are often asked to consult and help cardiologists with abandoned leads and pacemakers; for this reason, we report a case of a “forgotten” pacemaker.

Pacemaker-induced ventricular tachycardia and other manifestations of pacemaker malfunction associated with battery depletion were first described in the literature in 1965 [4]. Pacemaker-mediated tachycardias can interfere with the function of dual-chamber pacing systems and sensor-driven single-chamber rate-adaptive pacemakers. However, in rare cases, runaway pacing may induce potentially life-threatening tachycardias [5]. Reported mechanisms include sudden failure of electronic circuits, battery depletion, and circuit damage associated with electrical diathermy or therapeutic radiation [6, 7]. A failure of the transistor in the timing circuit has been reported, along with direct current leakage caused by failure of the pacemaker output capacitor [3, 4]. Such leakage has been proposed to cause polarization and increased myocardial resistance. Potential consequences include ventricular fibrillation or ventricular tachyarrhythmias, loss of capture with bradyarrhythmias, or asystole in pacemaker-dependent patients. Reported interventions for pacemaker-mediated tachycardia include magnetic inhibition, interrogation and reprogramming of the pulse generator, overdrive pacing, defibrillation, and emergent surgical procedures such as generator replacement or transection of pacing electrodes. While generator reprogramming and overstimulation have been suggested in the literature, they have shown limited success. Applying a magnet to the device can restore the baseline rhythm; however, if ineffective, urgent surgical intervention, such as generator replacement or lead transection, may be necessary [3]. If the patient is stimulation-dependent, disabling the malfunctioning system presents a new challenge. In the present case, the patient displayed spontaneous rhythm and effective pacing with a DDD transvenous pacemaker (TPC).

All modern cardiac implantable electronic devices incorporate an RPS protection circuit, preventing pacing above a preset rate (180–200 beats per minute [bpm]) [6]. It works by decreasing the pulse amplitude at high rates, concomitant with a loss of capture. RPS has also been described in patients with implantable cardioverter defibrillators [7]. Hermetic sealing is a crucial preventive measure in today’s world. Although RPS is rare and potentially severe, it can still occur in clinical practice. Therefore, we present a case illustrating the preventive management of an abandoned pacemaker system, notable for its 37-year pacing history, the presence of multiple abandoned systems, and the eventual surgical explantation of the original device as a preventive measure.

A 46-year-old male patient was admitted to our clinic for a definitive procedure with two implanted pacemakers (Figure 1 A, chest X-ray). The patient was diagnosed with an incomplete atrioventricular canal with a cleft in the anterior leaflet of the mitral valve at the age of 8 years. Surgical repair was performed on March 21, 1984. An unusual finding during the surgery was that the coronary sinus drained to the left side of the atrial septum. The cleft in the anterior mitral leaflet was closed with sutures, resulting in a competent mitral valve. The atrial septal defect was closed with a large patch. The early postoperative period was uneventful. On the 16th postoperative day, the patient developed a complete atrioventricular block followed by epicardial single chamber pacemaker (EPC) implantation (Siemens Elema Multi programmable pulse generator 668). In 2000, a TPC was implanted in VDD mode due to the high stimulation threshold of the EPC. In 2007, a new lead was implanted due to loss of capture with the abandoned old lead. In 2015, during elective replacement, an atrial lead and a DDD pacemaker (Vitatron E60A1 DR) were implanted due to sick sinus syndrome (chronotropic incompetence during exercise). On admission, the TPC was programmed to DDD 50 bpm, with ventricular pacing showing suitable sensing, threshold, and impedance parameters (Figure 2 A). Repeated lead tests were normal during exercise as well. During the hospital assessment, the patient’s own intrinsic rhythm was confirmed, with the DDD reprogrammed to VVI 30 bpm (Figure 2 B). The underlying rhythm was a complete AV block with escape beats at 36 bpm, accompanied by the warming-up phenomenon (Figure 2 C). The warming-up phenomenon occurs when the ventricular rhythm gradually increases in the presence of complete atrioventricular block (from HR below 30 to 34–36 bpm). The magnet application placed over the EPC generator location in the abdominal region revealed its activity with a magnet rate of 77 bpm (Figure 2 D). The estimated remaining longevity of the DDD EPC was 20 months. The decision was made to urgently perform a surgical explant of the EPC generator to prevent RPS. The generator reprogramming was impossible because the manufacturer programmer was unavailable for many years in most Polish centers. The explantation from the pocket in the straight abdominal sheath was performed on 9.12.2021. Local anesthesia (0.5% lidocaine) administered with sedation (midazolam and fentanyl i.v.) was used, and the incision was closed in layers (Figures 1 B, C). The postoperative period and wound healing were uneventful at the 6-month follow-up (Figure 1 D). This very late pacemaker removal was undertaken as a preventive measure against the potential complications of runaway pacemaker syndrome.

Figure 1

A – Chest X-ray, two PCs: epicardial and transvenous, multiple abandoned leads. B, C – Intraoperative pictures from the EPC explantation; D – wound healing at 6 months follow-up

EPC – epicardial single chamber pacemaker.

/f/fulltexts/KiTP/58747/KITP-23-3-58747-g001_min.jpg
Figure 2

A – Electrocardiogram on admission: DDD 50 bpm; B – transvenous pacemaker reprogrammed to VVI 30 bpm; C – patient’s underlying rhythm after several minutes: complete AV block with escape beats at 36 bpm, demonstrating the warming-up phenomenon; D – a magnet placed over the abdominal pocket for the EPC generator: magnetic rate of 77 bpm revealing ERI

/f/fulltexts/KiTP/58747/KITP-23-3-58747-g002_min.jpg

Runaway pacemaker syndrome could be avoided by timely generator replacement, especially in older pacemakers. Regular monitoring is essential for pacemaker patients to prevent battery depletion, with follow-up visits becoming more frequent as the device approaches its elective replacement period. If RPS occurs, the pulse generator must urgently be explanted. Early recognition of this medical emergency is crucial, as most patients present to the emergency department late, with syncopal or pre-syncopal symptoms.

Ethical approval

Not applicable.

Disclosures

The authors report no conflict of interest.

References

1 

Vijayaraman P, Vaidya K, Kim SG, Gross JN, Ferrick KJ, Palma E, Furman S, Fisher JD. Runaway pulse generator malfunction resulting from undetected battery depletion. Pacing Clin Electrophysiol 2002; 25: 220–222.

2 

Ortega DF, Sammartino MV, Pellegrino GM, Barja LD, Albina G, Segura EV, Balado R, Laino R, Giniger AG. Runaway pacemaker: a forgotten phenomenon? Europace 2005; 7: 592–597.

3 

Mickley H, Andersen C, Nielsen LH. Runaway pacemaker: a still existing complication and therapeutic guidelines. Clin Cardiol 1989; 12: 412–414.

4 

Wisoff BG, Gabor GE, Donoso E. Pacemaker-induced ventricular tachycardia. JAMA 1965; 192: 258–259.

5 

Vogelgesang D, Vogelgesang S. Pacemaker-induced ventricular tachycardia. Europace 2008; 10: 46–47.

6 

Böhm A, Hajdu L, Pintér A, Duray G, Préda I. Runaway pacemaker syndrome and intermittent nonoutput as manifestations of end of life of a VVI pacemaker. Pacing Clin Electrophysiol 2000; 23: 2143–2144.

7 

Carpenter CM, Galvin J, Guy M, McGovern BA. Runaway pacemaker in an implantable cardioverter defibrillator. J Cardiovasc Electrophysiol 1998; 9: 1008–1011.

Copyright: © 2026 Polish Society of Cardiothoracic Surgeons (Polskie Towarzystwo KardioTorakochirurgów) and the editors of the Polish Journal of Cardio-Thoracic Surgery (Kardiochirurgia i Torakochirurgia Polska). This is an Open Access article distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 International (CC BY-NC-SA 4.0) License (http://creativecommons.org/licenses/by-nc-sa/4.0/), allowing third parties to copy and redistribute the material in any medium or format and to remix, transform, and build upon the material, provided the original work is properly cited and states its license.
Share
without publication fees
without publication fees