Kardiochirurgia i Torakochirurgia Polska

Full text

3/2026 vol. 23
Original paper

Temporary stent-post cinching to facilitate minimally invasive implantation of the Avalus aortic bioprosthesis

  1. NH Rabindranath Tagore International Institute of Cardiac Sciences, Kolkata, India

Kardiochirurgia i Torakochirurgia Polska 2026; 23 (3): 220-222

Data publikacji online: 2026/10/12
Article file
Temporary stent-post.pdf

Introduction

Minimally invasive cardiac surgery (MICS) has gained popularity due to its reduced surgical trauma [1]. Advances in prosthetic valve design, including shorter struts and novel fixation mechanisms (e.g., Cor-knots), have facilitated MICS adoption. However, limited visualisation and the risk of leaflet damage during bioprosthetic valve implantation remain concerns. While traditional valves incorporate an external cinching mechanism to protect leaflets, the Avalus bioprosthesis lacks this feature.

Aim

We describe a simple technique to prevent leaflet damage during Avalus bioprosthesis implantation [2].

Material and methods

Surgical technique

Before the implantation of the Avalus bioprosthetic valve in the aortic position, a temporary artificial cinch is created on the operating table. A simple 5-0 polypropylene suture is passed around the tips of the stent posts sequentially, from one post to the next, until it returns to the first post. The suture is then tied sufficiently to bring the posts nearer each other, thereby creating a temporary cinch. Care should be taken not to completely distort the stent posts, as the cinch serves only as a protective mechanism during valve implantation. Then the valve is implanted as usual, either by continuous suturing or by interrupted suturing. After the implantation of the valve, the temporary cinch that is implanted is then removed by simply cutting the suture and removing it.

Discussion

Bioprosthetic valves are commonly supplied with mechanisms designed to protect the leaflets during implantation. Examples include the Inspiris Resilia aortic bioprosthesis (Edwards Lifesciences LLC, Irvine, CA, USA), EPIC (SJM, Minneapolis, USA), Magna Ease (Edwards, Lifesciences Corp, Irvine, CA), Mosaic bioprosthesis (Medtronic Inc, Minneapolis, MN), Intuity Elite (Edwards Lifesciences, Irvine, CA), Hancock II (Medtronic, Minneapolis, MN), and Daffodil Neo (Meril Life Sciences Pvt. Ltd, India). These mechanisms serve several purposes:

  1. Maintaining strut alignment: ensuring proper orientation of valve struts.

  2. Facilitating minimally invasive implantation: compressing the valve for easier insertion.

  3. Reducing the risk of leaflet damage: shielding leaflets from suture entanglement and trauma [3].

Most valves typically incorporate an acetyl polymer base ring, an Elgiloy stent frame, and leaflets positioned so as to minimize damage during valve deployment [4].

In contrast, the Avalus bioprosthesis utilises a dual-component frame design with a firm base constructed from polyetheretherketone (PEEK). The second component is a flexible trileaflet support of polyetheretherketone (PEEK) for the alpha-amino oleic acid (AOA) treated bovine pericardial leaflets. We have utilised the flexibility of PEEK in our technique, as described by Medtronic.

Unfortunately, the Avalus valve lacks a leaflet cinching mechanism and is instead supplied with a three-arm holder attached to the base ring near the nadir of each cusp. This leaves the stent posts exposed, exactly midway between two holder arms, with the leaflets potentially more susceptible to damage, particularly during minimally invasive surgery. Ideally, the three arms of the holder might have been attached to the stent posts themselves, potentially providing greater protection of the stent posts and leaflets during implantation. Instead, this design prominently exposes the delicate points, open to inadvertent damage.

Medtronic describes this design as having a low profile with minimised commissural posts. Regardless of the rationale for the design, the valve is currently available with this holder configuration, and measures to ensure safe implantation are therefore important. The response from Medtronic regarding our technique reassuring and did not raise concerns regarding the structural integrity of the product (Figure 1).

Figure 1

Avalus bioprosthetic valve with the holder showing the exposed stent posts

/f/fulltexts/KiTP/58767/KITP-23-3-58767-g001_min.jpg

We have evolved a simple workaround for the surgeon to address this issue on the operating table. We pass a polypropylene suture around the tip of each stent post, sequentially from one post to the next, until it returns to the first post, and then tie the knot just sufficiently to bring the posts closer together, thereby artificially creating a cinch. It is imperative that the knot be tightened only enough to create a cinch, producing an acute angle without completely collapsing or distorting the stent posts. This helps in moving the flexible stent posts from a right angle to an acute inward angle, improving exposure while simultaneously bridging the struts and preventing suture entanglement with the leaflet (Figures 2 and 3).

Figure 2

Polypropylene suture being passed through the stent post

/f/fulltexts/KiTP/58767/KITP-23-3-58767-g002_min.jpg
Figure 3

Avalus bioprosthetic valve after the cinching of the valve leaflets

/f/fulltexts/KiTP/58767/KITP-23-3-58767-g003_min.jpg

This temporary cinch mechanism can be easily created on the operating table. The holder can actually be removed with a little manipulation, even before the valve sutures are tied, provided that the cinch is not tightened excessively. Then the valve sutures are tied down with the leaflets still safely protected by the temporary cinch. Once the valve is seated and tied down, the cinch is easy to dismantle with a single cut of the temporary polypropylene suture (Figure 4).

Figure 4

Image showing suture entanglement around the stent post and the leaflet being protected by the cinch mechanism

/f/fulltexts/KiTP/58767/KITP-23-3-58767-g004_min.jpg

Our experience with this technique has been very satisfactory, and we feel it should be part of the surgical armamentarium available to cardiac surgeons.

Conclusions

Artificially cinching the stent posts during surgery may help prevent leaflet damage during implantation of the Avalus bioprosthesis. This technique is particularly useful in MICS and may also be considered during conventional median sternotomy. Our experience with this technique has been satisfactory, with no observed complications related to the technique.

Ethical approval

Compliant with HIPAA guidelines. Not needed as per institutional ethical guidelines.

Disclosures

The authors report no conflict of interest.

References

1 

Ling Y, Chen H, Zhu P, Li T, Xue B, Liu J. Minimally invasive valve surgery including patients of combined simultaneous surgery: a retrospective study. J Cardiothorac Surg 2023; 18: 266.

2 

Klautza RJM, Kappetein AP, Lange R, Dagenais F, Labrousse L, Bapat V, Moront M, Misfeld M, Zeng C, Sabik III JF; the PERIGON Investigators. Safety, effectiveness, and haemodynamic performance of a new stented aortic valve bioprosthesis. J Thorac Cardiovasc Surg 2024; 52: 425–431.

3 

Durko AP, Head SJ, Pibarot P, Atluri P, Bapat V, Cameron DE, Casselman FPA, Chen EP, Dahle G, Ebels T, Elefteriades JA, Lancellotti P, Prager RL, Rosenhek R, Speir A, Stijnen M, Tasca G, Yoganathan A, Walther T, De Paulis R; EACTS–STS–AATS Valve Labelling Task Force. Characteristics of surgical prosthetic heart valves and problems around labeling: a document from the European Association for Cardio-Thoracic Surgery (EACTS)–The Society of Thoracic Surgeons (STS)–American Association for Thoracic Surgery (AATS) Valve Labelling Task Force. J Thorac Cardiovasc Surg 2019; 158: 1041–1054.

4 

Kostyunin AE, Yuzhalin AE, Rezvova MA, Ovcharenko EA, Glushkova TV, Kutikhin AG. Degeneration of bioprosthetic heart valves: update 2020. J Am Heart Assoc 2020; 9: e018506.

Copyright: © 2026 Polish Society of Cardiothoracic Surgeons (Polskie Towarzystwo KardioTorakochirurgów) and the editors of the Polish Journal of Cardio-Thoracic Surgery (Kardiochirurgia i Torakochirurgia Polska). This is an Open Access article distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 International (CC BY-NC-SA 4.0) License (http://creativecommons.org/licenses/by-nc-sa/4.0/), allowing third parties to copy and redistribute the material in any medium or format and to remix, transform, and build upon the material, provided the original work is properly cited and states its license.