Introduction
Minimally invasive cardiac surgery (MICS) has gained popularity due to its reduced surgical trauma [1]. Advances in prosthetic valve design, including shorter struts and novel fixation mechanisms (e.g., Cor-knots), have facilitated MICS adoption. However, limited visualisation and the risk of leaflet damage during bioprosthetic valve implantation remain concerns. While traditional valves incorporate an external cinching mechanism to protect leaflets, the Avalus bioprosthesis lacks this feature.
Aim
We describe a simple technique to prevent leaflet damage during Avalus bioprosthesis implantation [2].
Material and methods
Surgical technique
Before the implantation of the Avalus bioprosthetic valve in the aortic position, a temporary artificial cinch is created on the operating table. A simple 5-0 polypropylene suture is passed around the tips of the stent posts sequentially, from one post to the next, until it returns to the first post. The suture is then tied sufficiently to bring the posts nearer each other, thereby creating a temporary cinch. Care should be taken not to completely distort the stent posts, as the cinch serves only as a protective mechanism during valve implantation. Then the valve is implanted as usual, either by continuous suturing or by interrupted suturing. After the implantation of the valve, the temporary cinch that is implanted is then removed by simply cutting the suture and removing it.
Discussion
Bioprosthetic valves are commonly supplied with mechanisms designed to protect the leaflets during implantation. Examples include the Inspiris Resilia aortic bioprosthesis (Edwards Lifesciences LLC, Irvine, CA, USA), EPIC (SJM, Minneapolis, USA), Magna Ease (Edwards, Lifesciences Corp, Irvine, CA), Mosaic bioprosthesis (Medtronic Inc, Minneapolis, MN), Intuity Elite (Edwards Lifesciences, Irvine, CA), Hancock II (Medtronic, Minneapolis, MN), and Daffodil Neo (Meril Life Sciences Pvt. Ltd, India). These mechanisms serve several purposes:
Maintaining strut alignment: ensuring proper orientation of valve struts.
Facilitating minimally invasive implantation: compressing the valve for easier insertion.
Reducing the risk of leaflet damage: shielding leaflets from suture entanglement and trauma [3].
Most valves typically incorporate an acetyl polymer base ring, an Elgiloy stent frame, and leaflets positioned so as to minimize damage during valve deployment [4].
In contrast, the Avalus bioprosthesis utilises a dual-component frame design with a firm base constructed from polyetheretherketone (PEEK). The second component is a flexible trileaflet support of polyetheretherketone (PEEK) for the alpha-amino oleic acid (AOA) treated bovine pericardial leaflets. We have utilised the flexibility of PEEK in our technique, as described by Medtronic.
Unfortunately, the Avalus valve lacks a leaflet cinching mechanism and is instead supplied with a three-arm holder attached to the base ring near the nadir of each cusp. This leaves the stent posts exposed, exactly midway between two holder arms, with the leaflets potentially more susceptible to damage, particularly during minimally invasive surgery. Ideally, the three arms of the holder might have been attached to the stent posts themselves, potentially providing greater protection of the stent posts and leaflets during implantation. Instead, this design prominently exposes the delicate points, open to inadvertent damage.
Medtronic describes this design as having a low profile with minimised commissural posts. Regardless of the rationale for the design, the valve is currently available with this holder configuration, and measures to ensure safe implantation are therefore important. The response from Medtronic regarding our technique reassuring and did not raise concerns regarding the structural integrity of the product (Figure 1).
We have evolved a simple workaround for the surgeon to address this issue on the operating table. We pass a polypropylene suture around the tip of each stent post, sequentially from one post to the next, until it returns to the first post, and then tie the knot just sufficiently to bring the posts closer together, thereby artificially creating a cinch. It is imperative that the knot be tightened only enough to create a cinch, producing an acute angle without completely collapsing or distorting the stent posts. This helps in moving the flexible stent posts from a right angle to an acute inward angle, improving exposure while simultaneously bridging the struts and preventing suture entanglement with the leaflet (Figures 2 and 3).
This temporary cinch mechanism can be easily created on the operating table. The holder can actually be removed with a little manipulation, even before the valve sutures are tied, provided that the cinch is not tightened excessively. Then the valve sutures are tied down with the leaflets still safely protected by the temporary cinch. Once the valve is seated and tied down, the cinch is easy to dismantle with a single cut of the temporary polypropylene suture (Figure 4).
Figure 4
Image showing suture entanglement around the stent post and the leaflet being protected by the cinch mechanism

Our experience with this technique has been very satisfactory, and we feel it should be part of the surgical armamentarium available to cardiac surgeons.
Conclusions
Artificially cinching the stent posts during surgery may help prevent leaflet damage during implantation of the Avalus bioprosthesis. This technique is particularly useful in MICS and may also be considered during conventional median sternotomy. Our experience with this technique has been satisfactory, with no observed complications related to the technique.