Abstract
Ultreon 3.0 and physiological coronary assessment: initial single-center Polish experience
Clinical Department of Cardiology and Cardiovascular Interventions, University Hospital, Krakow, Poland
Second Department of Cardiology, Institute of Cardiology, Jagiellonian University Medical College, Krakow, Poland
Department of Interventional Cardiology, Institute of Cardiology, Jagiellonian University Medical College, Krakow, Poland
Clinical Department of Interventional Cardiology, St. John Paul II Hospital, Krakow, Poland
Royal Victoria Hospital, McGill University Health Center, Montreal, QC, Canada
Adv Interv Cardiol 2026; 22, 2 (84): 217–223
Introduction
Optical coherence tomography–derived fractional flow reserve (FFR-OCT), computed as virtual flow reserve, combines anatomical and functional lesion assessment without pressure-wire manipulation or pharmacological hyperemia. Its output may depend on accurate lumen delineation and side-branch recognition.
Aim
To assess the agreement between operator-reviewed FFR-OCT and wire-based FFR in routine practice, and the effect of manual segmentation correction.
Material and methods
We analyzed 14 consecutive patients (19 vessels) undergoing wire-based FFR followed by OCT-guided percutaneous coronary intervention (PCI). OCT pullbacks were processed with Ultreon 3.0 software; automated FFR-OCT values were reviewed by an experienced operator with manual correction of lumen and side-branch segmentation. Comparison used Pearson’s correlation, Bland–Altman analysis, and concordance in ischemia classification at the 0.80 threshold.
Results
Wire-based FFR was 0.79 ±0.10. In 13 vessels with both values, automated FFR-OCT increased from 0.68 ±0.16 to 0.75 ±0.11 after operator review (mean paired difference +0.064; Wilcoxon p = 0.016). Across 19 vessels, operator-reviewed FFR-OCT (0.76 ±0.10) correlated moderately with wire-based FFR (r = 0.614, 95% CI: 0.222–0.835; p = 0.005; bias +0.025, 95% limits of agreement −0.153 to +0.203). Classification was concordant in 15 of 19 vessels; 4 were discordant (21.1%, 95% CI: 6.1–45.6%).
Conclusions
Manual review of segmentation substantially changed the automated FFR-OCT result. Although operator-reviewed FFR-OCT was associated with wire-based FFR, agreement near the 0.80 threshold was imperfect. FFR-OCT may aid PCI planning but does not yet appear interchangeable with pressure-wire physiology. Larger prospective studies with prespecified, blinded image analysis are warranted.
Keywords
coronary physiology, fractional flow reserve, virtual flow reserve, optical coherence tomography, Ultreon 30