Introduction

The growing use of mechanical circulatory support (MCS) devices in pediatric patients has expanded therapeutic options for end-stage heart failure. Technologies such as microaxial flow pumps and durable left ventricular assist devices (LVADs) now provide life-saving therapeutic options, most commonly as a bridge to heart transplantation [1, 2]. While single-device strategies are well established, staged approaches involving multiple devices remain less common but may be essential in cases of profound hemodynamic instability.

This report presents a case of a pediatric patient with end-stage cardiomyopathy and cardiogenic shock successfully managed using a staged, triple-bridge MCS strategy. The approach included initial stabilization with a microaxial flow pump, transition to durable LVAD support, and temporary right ventricular assist device (tRVAD) implantation to address perioperative right ventricular failure. The case highlights both the feasibility and complexity of this strategy, emphasizing technical considerations and clinical decision-making.

Initial presentation and Impella CP support

A 10-year-old female (40 kg, body surface area 1.30 m2) was admitted with acute decompensated dilated cardiomyopathy and cardiogenic shock (SCAI stage C, Pedimacs profile 1). This was her first presentation of heart failure, with no prior medical history but a positive family history of cardiomyopathy. She initially presented with dyspnea, chest pain, and vomiting and required escalating inotropic support, including noradrenaline, levosimendan, and adrenaline.

Echocardiography demonstrated severe left ventricular dysfunction (ejection fraction 10%), significant left ventricular dilation, severe mitral regurgitation, and mildly impaired right ventricular function. Due to progressive hemodynamic deterioration, urgent implantation of a percutaneous microaxial flow pump (Impella CP) was performed.

The device was inserted via the right femoral artery under ultrasound and fluoroscopic guidance. Proper positioning – with the inlet in the left ventricle and outlet in the ascending aorta – was confirmed using transesophageal echocardiography. Gradual escalation of device support allowed effective unloading of the left ventricle while maintaining systemic perfusion.

Anticoagulation was maintained with unfractionated heparin targeting an activated clotting time of 160–180 s. Continuous monitoring included hemodynamic parameters, echocardiographic assessment, and laboratory markers of hemolysis. No major complications such as bleeding, limb ischemia, or device-related injury occurred.

Before device implantation, the patient exhibited signs of severe end-organ hypoperfusion, including elevated lactate, renal dysfunction (creatinine 87 µmol/l), and hepatic impairment (bilirubin 22.6 µmol/l, INR 1.52). Following initiation of support, rapid improvement was observed. Lactate levels normalized, renal function improved significantly, and hepatic parameters returned toward normal. These findings indicated effective restoration of systemic perfusion.

Such biochemical recovery is widely considered a marker of reversible shock under temporary MCS. In contrast, persistent metabolic derangement or worsening organ dysfunction would suggest the need for escalation to durable support. In this case, favorable trends justified continuation of temporary support as a bridge strategy.

Despite hemodynamic stabilization and reduction of inotropic requirements, there was no evidence of intrinsic myocardial recovery. Attempts to reduce device support resulted in hypotension and worsening perfusion. Persistent left ventricular dysfunction, elevated filling pressures, and ventricular dilation confirmed the need for transition to durable LVAD therapy.

Transition to durable LVAD and tRVAD support

After 7 days of temporary support, the patient underwent implantation of a durable LVAD (HeartMate 3) via median sternotomy. The inflow cannula was placed at the left ventricular apex, and the outflow graft was anastomosed to the ascending aorta.

However, attempts to wean from cardiopulmonary bypass were unsuccessful due to severe right ventricular failure. Intraoperative findings included right ventricular dilation, poor contractility, severe tricuspid regurgitation, and septal shift. These findings were associated with inadequate LVAD flow and elevated central venous pressure.

To address this, a temporary right ventricular assist device (CentriMag) was implanted. The inflow cannula was inserted via the femoral vein, and the outflow cannula was connected to the pulmonary artery. This configuration provided effective right-sided circulatory support while maintaining LVAD function.

Postoperatively, LVAD flows were maintained above 3.0 l/min, with speeds adjusted to ensure optimal septal positioning and avoid suction events. RVAD flow ranged between 3.8 and 3.9 l/min. Both devices were carefully managed as a coupled system, requiring continuous adjustment to prevent hemodynamic imbalance.

Right ventricular recovery and tRVAD explantation

Assessment of right ventricular recovery involved a combination of clinical, hemodynamic, laboratory, and echocardiographic parameters. Weaning of tRVAD support began 48 hours after implantation once initial signs of recovery were observed.

The weaning process included gradual reduction of RVAD flow in controlled increments, with close monitoring of hemodynamics and echocardiographic findings. A low-flow trial (1.0 l/min) was maintained for up to 24 hours to assess the patient’s ability to tolerate minimal support.

Criteria indicating recovery included stable mean arterial pressure without vasopressor escalation, central venous pressure ≤ 12 mm Hg, improved right ventricular contractility, reduced chamber dilation, and stable LVAD performance. Additionally, improvement in renal and hepatic function and decreasing lactate levels supported clinical recovery.

On postoperative day 7, the patient underwent successful explantation of the tRVAD. Throughout this process, careful coordination between LVAD and RVAD settings was essential to avoid complications such as septal shift, pulmonary congestion, or ventricular suction events. Perioperative and postoperative data are summarized in Table 1.

Table 1

Intraoperative and postoperative data

ParameterData
Cardiopulmonary bypass duration [minutes]152
Intensive care unit stay [days]47
Postoperative intravenous inotropes; durationAdrenaline, noradrenaline, milrinone, levosimendan; 22 days
Hospital stay [days]50
Early postoperative complicationsPleural effusion
Pulmonary hypertension treatmentiNO, sildenafil
Follow-up duration [months]4

Hospital course and transition to transplantation

Histological examination revealed evidence of prior myocardial injury, including organized infarction and mural thrombus formation. The patient required prolonged intensive care, with an ICU stay of 35 days and total hospitalization of 47 days. Inotropic support was necessary for 22 days, along with pulmonary vasodilator therapy.

The patient was discharged home in stable condition on LVAD support with appropriate anticoagulation. During this period, no device-related complications occurred, including thrombosis, stroke, or infection.

After approximately 1 month of durable LVAD support, the patient underwent successful orthotopic heart transplantation. The perioperative course was uneventful, with rapid recovery of graft and end-organ function. She was discharged in stable condition and remained clinically well at 3-month follow-up.

Table 2 illustrates the timeline of clinical events. Informed consent for publication was obtained from the patient’s legal guardian.

Table 2

Timeline of clinical events

DateProcedure
21.8.2025Impella CP implantation
28.8.2025HM3 implantation, intraoperative tRVAD implantation
4.9.2025tRVAD explantation
3.10.2025Discharged home
10.10.2025Heart transplantation
31.10.2025Discharged home after heart transplantation

[i] HM3 – HeartMate 3, tRVAD – temporary right ventricular assist device.

Discussion

This case demonstrates the feasibility of a staged, triple-bridge MCS strategy in a pediatric patient with severe cardiogenic shock. Initial stabilization with a microaxial flow pump was associated with rapid restoration of systemic perfusion and improvement in end-organ dysfunction. Transition to durable LVAD support provided long-term circulatory assistance, while temporary RVAD support addressed acute right ventricular failure [3].

The increasing use of microaxial pumps in pediatric populations has been supported by recent clinical data [1, 4], with studies reporting high rates of successful bridging to transplantation and favorable survival outcomes [5, 6]. Similarly, durable LVADs are increasingly recognized as an important option in children, enabling extended support while awaiting transplantation [2, 7–9].

Right ventricular failure remains a major challenge following LVAD implantation. This case highlights the importance of early recognition and timely intervention with temporary RV support. Successful recovery and explantation may depend on careful patient selection, systematic weaning protocols, and comprehensive monitoring [2].

The decision to employ a staged, triple-bridge strategy was influenced by the inability to secure a donor organ promptly. This reflects a broader challenge in pediatric transplantation, particularly in regions with limited donor availability. In such settings, advanced MCS strategies may be essential to bridge patients safely to transplantation.

Conclusions

This report illustrates the feasibility of a staged, triple-bridge mechanical circulatory support strategy in a pediatric patient with severe cardiogenic shock. The combination of temporary microaxial support, durable LVAD therapy, and temporary RVAD support enabled stabilization, recovery of organ function, and eventual successful heart transplantation.

The case underscores the importance of individualized treatment strategies, multidisciplinary management, and adaptability in the use of modern MCS technologies. As experience with these devices continues to grow, such approaches may become increasingly important in managing complex pediatric heart failure cases.

Ethical approval

Not applicable.

Disclosures

The authors report no conflict of interest.

References

1 

Mehdizadeh-Shrifi A, Ahmed HF, Chappel G, Faateh M, Perry T, Lorts A, Morales DLS, Ashfaq A. The increasing utilization of the Impella device as abridge-to-transplantion in pediatric heart centers across the United States. World J Pediatr Congeit Heart Surg 2025; 16: 769–775.

2 

O´Connor MJ, Shezad M, Ahmed H, Amdani S, Auerbach SR, Bearl DW, Butto A, Byrnes JW, Conway J, Dykes JC, Glass L, Lantz J, Law S, Mongé MC, Morales DLS, Parent JJ, Peng DM, Ploutz MS, Puri K, Shugh S, Shwaish NS, Vander-Pluym CJ, Wilkens S, Wright L, Zinn MD, Lorts A. Expanding use of the Heart-Mate 3 ventricular assist device in pediatric and adult patients within the advanced cardiac therapies improving outcomes network (ACTION). J Heart Lung Transplant 2023; 42: 1546–1556.

3 

Powel TR. Early intraoperative detection and management of right ventricular failure after left ventricular assist device implantation. Tex Heart Inst J 2023 50: e238169.

4 

Edelson JB, Amdani S, Rosenthal DN, Kroslowtz B, Morray B, Law S, Joong A, Wilkens S, Peng DM, Ploutz M, Murray J, Lantz J, Esch JJ, Iqbal M, Lancaster T, Vander-Pluym C, Jeewa A, Shugh S, Absi M, Shezad M, O‘Connor MJ, Lorts A, Tume S. Multi-institutional outcomes of Impella use in pediatric patients: a brief communication from the ACTION Network J Heart Lung Transplant 2025; 44: 1668–1671.

5 

Alsoufi B, Kozik D, Oelkers B, Wilkens S, Sparks J, Trivendi J. Efficacy of Impella microaxial left ventricular assist device as a bridge to transplant in children with end-stage heart disease. Interdiscip Cardiovasc Thorac Surg 2025; 40: ivaf249.

6 

Kushnerus K, Potapv E, Lanmuller P, Starck C, Cho MY, Photiadis J. First results of the Deutsches Herzzntrum der Charite paediatric impella cohort. Eur J Cardiothorac Surg 2025; 67: ezaf019.

7 

Rohde S, van Puyvelde J, Veen KM, Schweiger M, Bierman D, Amodeo A, Martens T, Damman K, Gollmann-Tepeköylü C, Hulman M, Iacovoni A, Krämer US, Loforte A, Napoleone CP, Nemec P, Netuka I, Özbaran M, Polo L, Pya Y, Ramjankhan F, Sandica E, Sliwka J, Stiller B, Kadner A, Franceschini A, Thiruchelvam T, Zimpfer D, Berger F, Davies B, Dashkevich A, Stark C, Meyns B, de By TMMH, Miera O. The European registry for patients with mechanical circulatory support (EUROMACS): fourth Paediatric EUROMACS (Paedi-EUROMACS) report. Eur J Cardiothorac Surg 2024; 66: ezae276.

8 

Adachi I, Peng DM, Hollander SA, Simpson KE, Davies RR, Jacobs JP, Vander-Pluym CJ, Fynn-Thompson F, Wells DA, Law SP, Amdani S, Cantor R, Koehl D, Kirklin JK, Morales DLS, Rossano JW; Pedimacs Investigators. Sixth annual pediatric interagency registry for mechanical circulatory support (Pedimacs) report: the society of thoracic surgeons Pedimacs annual report. Ann Thorac Surg 2023; 115: 1098–1108.

9 

Bilgili A, Pendarvis J, Sharaf OM, Peek GJ, Bleiweis MS, Jacobs JP. Analysis of UNOS: ventricular assist device as bridge-to-transplant in paediatric patients with congenital heart disease. Eur J Cardiothorac Surg 2025; 67. doi:10.1093/ejcts/ezaf290.

Copyright: © 2026 Polish Society of Cardiothoracic Surgeons (Polskie Towarzystwo KardioTorakochirurgów) and the editors of the Polish Journal of Cardio-Thoracic Surgery (Kardiochirurgia i Torakochirurgia Polska). This is an Open Access article distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 International (CC BY-NC-SA 4.0) License (http://creativecommons.org/licenses/by-nc-sa/4.0/), allowing third parties to copy and redistribute the material in any medium or format and to remix, transform, and build upon the material, provided the original work is properly cited and states its license.
Share
without publication fees
without publication fees