Methods and Techniques
Stent loss in the radial artery – surgical vs. interventional approach – report of two cases

Artur Baszko
,
Wojciech Telec
,
Eryk Naumowicz
,
Tomasz Siminiak
,
Piotr Kałmucki

Postep Kardiol Inter 2015; 11, 1 (39): 50–54
Online publish date: 2015/03/06
Article file
Get citation
 
 

Case reports

Case 1

A 58-year-old patient with a history of hypertension, hypercholesterolaemia, previous left circumflex artery (LCx) angioplasty (2007), and an attempt of percutaneous coronary intervention (PCI) to the right coronary artery (RCA) (2008) presented with progressive unstable angina despite optimal medical treatment. In the emergency room the patient presented with ST segment depression on electrocardiogram (ECG), but cTnT was negative. After preparing the patient with aspirin 300 mg and clopidogrel 600 mg transradial coronary angiography (6 Fr Sheath, Terumo) revealed noncritical, multisegmental lesions in the left anterior descending artery (LAD), occluded distal LCx, and long critical, eccentric narrowing in the mid-to-distal part of the RCA with hardly any visible calcifications.
The procedure of PCI of the RCA was attempted. A Whisper ES (Whisper ES, Abbott) guidewire was inserted and several predilations were performed using 2.0 × 20 mm and 2.5 × 20 mm angioplasty balloons (Sprinter Legend, Medtronic). The drug eluting stent did not cross the distal calcified lesion (Xience Prime 3.5 × 28 mm, Abbott). During the stent pullback to the guiding catheter it slipped out leaving an approximately 2–3 mm margin covering the distal part of the balloon. The stent could not be advanced again, and an attempt to remove it was made. In order to do so, the balloon was inflated to 6 atm to fixate the stent. The stent, along with the inflated balloon and catheter, were pulled back together into the sheath (Figures 1–2). As we could not get the stent into the sheath the entire assembly was removed en bloc, unfortunately leaving the stent in the radial artery in the wrist region. Light compression was applied to stop bleeding, and under X-ray the location of the stent was confirmed at the level of artery puncture.
A vascular surgeon was asked for consultation and it was decided to perform a radial artery cut-down to surgically remove the stent (Figures 3–5). This was done under local anaesthesia without complications.
As the patient experienced chest discomfort due to RCA dissection the procedure was continued using the transfemoral approach. The procedure required using two extra support guidewires, non-compliant balloons, and stronger support with an AL1 guiding catheter. Finally the stent was deployed (Xience Prime 3.0 × 28 mm) and additional post-dilation with a 3.5 × 15 mm balloon (NC Sprinter RX, Medtronic) resulted in a very good angiographic effect. During hospitalisation there was a slight oedema of the operated hand, but pulses on both radial and ulnar arteries were maintained. A further 18-month follow-up was uneventful.

Case 2

A 53-year-old patient presented with chest pain and ST segment elevation consistent with inferior infarction. Previous history revealed hypertension treated with ramipril. The patient received aspirin 300 mg, clopidogrel 600 mg, and 5000 U of unfractionated heparin in the emergency room and was sent for urgent angiography. The procedure was performed transradially (6 Fr sheath, Terrumo) and revealed 65% narrowing in mid RCA suitable for “direct stenting”.
The Whisper MS guidewire was inserted and a stent (3.0 × 15 mm Xience Prime, Abbott) was positioned (Figure 6). During the attempt to deploy the stent, a pressure rise on the pump pressure gauge was not observed. Close inspection revealed a broken three-way cock connecting the pump with the balloon catheter. The connector was replaced. During this manoeuvre the stent was moved proximally to the lesion. An attempt to push it across the lesion failed, and an attempt to pull it back into the guiding catheter was performed. It was also impossible probably due to a slight but invisible inflation of the balloon during the first attempt to deploy the stent. On the second attempt the stent went into the guiding catheter. The balloon was partially inflated (6 atm) to press the stent against the catheter wall. Thereafter the catheter, the trapped stent, and the wire were removed slowly towards the sheath (Figure 7). After completing the pull back, there was no stent remaining in the catheter. Inspection under X-ray showed that the stent was lost in the proximal segments of the radial artery. The stent was rewired and deployed in situ using 3.5 × 15 mm and 4.0 × 15 mm balloons (Figure 8). Because a good result was obtained, the PCI was continued using the same approach. Introducing the 6 Fr guiding catheter led to stent distortion (Figures 9–10), and another 4.0 × 15 balloon was used to reshape it properly (Figure 11). After this, the guiding catheter was replaced with a 5 Fr catheter (Figure 12) and the procedure was completed without further complications. The next 14 months of patient follow-up were uneventful.

Discussion

Stent loss can be related to several factors such as: inserting while the balloon is exposed to negative pressure, inserting the stent through a previously deployed stent, advancing the stent through calcified, tortuous lesions or during withdrawal of the stent to the guiding catheter while the proximal struts are damaged [2, 3]. An artificial body left in the vascular system may lead to thrombosis, but in most sites it is benign unless the stent is within the central nervous system, heart, or arteries supplying vital organs. Embolisation of stents in coronary arteries has a hazardous potential for the patient and should be avoided due to high risk of complications [6].
The procedure performed through the radial artery creates some challenges, but it also offers some possibilities to deal with lost stents. We have presented two cases of stent loss. In the first case it was due to serious underestimation of calcifications, and in the second after an attempt to withdraw a partially inflated stent. Both stents were still within the coronary artery. Deploying the lost stent within the coronary artery was one option, which was proven to be easy and safe, and it did not disrupt the completion of the case. We decided to remove stents into the catheter because part of the balloon was still within the stent. The partially inflated balloon was used to anchor the stent to the wall of the catheter. However, in both cases stents were lost within the radial artery. The difference in stent position led to different approaches. In the first situation there was significant resistance at the entry to the sheath, and finally we attempted to remove the entire assembly (sheath, catheter, balloon, and stent) en block. The stent, however, slipped into the radial artery at the site of puncture, which was confirmed under fluoroscopy. This was an optimal situation for surgical intervention, which was performed immediately. In a similar case reported recently, when radial cut-down stent retrieval was performed, the PCI was postponed for several days [5]. In our case, the patient had chest pain due to the dissection of the artery. This is why we decided to perform radial artery cut-down on heparin and then continue with angioplasty. The patient was on a triple therapy (aspirin, clopidogrel, and unfractionated heparin), and this did not cause any post-surgical complications. This is in line with the observation of pacemaker patients who were on dual or triple therapy during surgery [7].
In the second case the stent was lost very high in the forearm – this location was not ideal for surgical intervention. We decided to deploy the stent with 3.5 and 4.0 mm balloons, and the procedure was continued by passing the stent with the use of a guidewire. Another useful technique that could be used in this case was a snare and guiding catheter, as described by Deftereos et al. [8]. In our case the stent was initially inflated in the proximal part, which could make the stent too large to insert into the guiding catheter.

Conclusions

In our opinion, deploying a lost stent within the coronary artery in a healthy segment should not be performed as long as a chance to remove it exists. Having the stent fixed with the tip of the balloon gives the opportunity to remove it to the radial artery. Surgical extraction or deployment of the stent in the radial artery was successful in both cases, which enabled us to complete the case, and it had no negative impact during long-term follow-up.

Conflict of interest

The authors declare no conflict of interest.

References

1. Dunning DW, Kahn JK, O’Neill WW. The long-term consequences of lost intracoronary stents. J Interv Cardiol 2002; 15: 345-8.
2. Brialakis ES, Best PJ, Elesber AA, et al. Incidence, retrieval methods, and outcomes of stent lost during percutaneous coronary intervention: a large single-center experience. Catheter Cardiovasc Interv 2005; 55: 333-40.
3. Iturbe JM, Abdel-karim AR, Papayannis A, et al. Frequency, treatment, and consequences of device loss and entrapment in contemporary percutaneous coronary interventions. J Invasive Cardiol 2012; 24: 215-21.
4. Ahmar W, Malaiapan Y, Meredith IT. Transradial retrieval of dislodged stent from the left coronary artery. J Invasive Cardiol 2008; 20: 545-7.
5. Cha KS. Surgical retrieval of dislodgeds stent during transradial coronary intervention. J Invasive Cardiol 2012; 24: E179-81.
6. Eggerbrecht H, Haude M, von Birgelen C, et al. Nonsurgical retrival of embolizes coronary artery stents. Catheter Cardiovasc Interv 2000; 51: 432-40.
7. Ghanbari H, Phard WS, Al-Ameri H, et al. Meta-analysis of safety and efficacy of uninterrupted warfarin compared to heparin-based bridging therapy during implantation of cardiac rhythm devices. Am J Cardiol 2012; 110: 1482-8.
8. Deftereos S, Raisakis K, Giannopoulos G, et al. Successflul retrieval of coronary stent dislodged in the brachial artery by means of improvised snare and guiding catheter. Int J Angiol 2011; 20: 55-8.
Copyright: © 2015 Termedia Sp. z o. o. This is an Open Access article distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 International (CC BY-NC-SA 4.0) License (http://creativecommons.org/licenses/by-nc-sa/4.0/), allowing third parties to copy and redistribute the material in any medium or format and to remix, transform, and build upon the material, provided the original work is properly cited and states its license.
Quick links
© 2024 Termedia Sp. z o.o.
Developed by Bentus.