eISSN: 2081-2841
ISSN: 1689-832X
Journal of Contemporary Brachytherapy
Current Issue Archive Supplements Articles in Press Journal Information Aims and Scope Editorial Office Editorial Board Register as Author Register as Reviewer Instructions for Authors Abstracting and indexing Subscription Advertising Information Links
Editorial System
Submit your Manuscript
SCImago Journal & Country Rank

5/2023
vol. 15
 
Share:
Share:
Original paper

Analysis of source dwell position during treatment in brachytherapy using CT scout images

Nozomi Nakajima
1, 2
,
Tomonori Isobe
1
,
Yoshinobu Furuyama
3
,
Tetsuya Tomita
4
,
Daisuke Kobayashi
4
,
Yutaro Mori
1
,
Hideyuki Takei
5
,
Yoshiaki Nagai
6
,
Yasuwo Ide
6
,
Keiko Nemoto Murofushi
7
,
Masaru Nakajima
2
,
Yasuo Yoshioka
2

1.
Graduate School of Comprehensive Human Sciences, University of Tsukuba, Ibaraki, Japan
2.
Radiation Oncology Department, Cancer Institute Hospital, Japanese Foundation for Cancer Research, Koto-ku, Tokyo, Japan
3.
Department of Radiation Oncology, Ibaraki Prefectural Central Hospital, Japan
4.
Department of Radiology, University of Tsukuba Hospital, Japan
5.
Quantum Life and Medical Science Directorate, National Institutes for Quantum Science and Technology, Japan
6.
Faculty of Health Science, Tsukuba International University, Japan
7.
Department of Radiology, Tokyo Metropolitan Cancer and Infectious Diseases Center Komagome Hospital, Japan
J Contemp Brachytherapy 2023; 15, 5: 357–364
Online publish date: 2023/10/30
Article file
- Analysis of source.pdf  [0.60 MB]
Get citation
 
 

Purpose

Radiation therapy is an important part of cancer treatment combined with chemotherapy and surgery. Radiation therapy can be broadly classified into two types, such as brachytherapy and external radiation. High-dose-rate (HDR) brachytherapy utilizes radiation emitted from a capsule (hereafter referred to as ‘source’) containing a radioisotope, such as iridium-192 (192Ir) or cobalt-60 (60Co). A tubular instrument, called an ‘applicator’ or ‘catheter’, is inserted into the patient providing a route, through which the source is delivered to the target, and the target is irradiated by the source. The extent and intensity of irradiation are determined by examining dwell time and position of the source using a radiotherapy planning system. A dose per fraction is defined to be approximately 5-6 Gy at a point A, as defined in the International Commission on Radiation Units report 38 [1]; a dose per fraction in brachytherapy is higher than that of a typical external radiation. Although higher doses are more effective in treating tumors, brachytherapy involves a higher risk of inaccurate irradiation to the patient. Methods to prevent radiation therapy errors have been developed for external radiation [2, 3]. However, these methods have not been established for brachytherapy, and their implementation depends on the efforts of each facility [4]. The International Commission on Radiological Protection (ICRP) Publication 97 reported the actual source dwell at a position that differ from the treatment plan, incidents in which the source is stuck to the applicator, and incidents due to using wrong length of catheters [5]. To detect these events, the Nuclear Regulatory Commission highlighted the importance of verifying treatment plans that involve irradiation [6]. Several studies have been conducted to detect the source dwell position during irradiation [7-18]. However, these methods require special equipment and instruments, and can be challenging to utilize. In addition, these methods analyze the relative source position; therefore, a mispositioned applicator may not be correctly detected. The purpose of the present study was to construct a system for analyzing the source position during irradiation using computed tomography (CT) scout images. The novelty of this study was that the system utilized CT images for image-guided brachytherapy, and is highly versatile enabling analysis of absolute coordinates.

Material and methods

Using a CT scout image, dwell position coordinates of the source during irradiation were analyzed using an in-house MATLAB program. A scout image is a 2-dimensional image acquired to determine an accurate scan volume and scan location [19]. In the acquisition of CT scout images under different imaging conditions section, we examined whether the source could be identified in CT scout images that were acquired though source dwelled in the applicator under different imaging conditions and phantom positions. In the source dwell position detection program section, we analyzed source dwell position from the acquired CT images using an in-house MATLAB program, and compared them with source dwell positions in treatment plans.

Acquisition of ct scout images under different imaging conditions

Computed tomography scout images were acquired while the source dwelled in the applicator. Figure 1 shows the standard arrangement of the applicator. A 15° tandem tube from the Fletcher Williamson Asian Pacific applicator set (Elekta, Sweden) was inserted into water-filled phantom (RT-3000, R-TECH, Japan) and fixed with an applicator clamp. This study was considered to be implemented into clinical treatments. The applicator in the water phantom simulated a patient under treatment. Dotted lines in two illustrations indicated a CT gantry center. The tandem tube tip was positioned in the center of CT gantry. Positioning accuracy of the CT gantry was ±1 mm.

Fig. 1

Installation of water phantom and tandem tube for a standard arrangement. A) An overall view of the set-up. B, C) The po -sition of tandem applicator in the CT gantry. Dotted lines indicate the CT gantry center

/f/fulltexts/JCB/51750/JCB-15-51750-g001_min.jpg

CT system used was SOMATOM Definition AS Open (Siemens Healthineers, Germany) that allowed only the gantry movement during imaging. Slice thickness and reconstructed slice thickness were both 0.6 mm, and tube voltage was set to 120 kVp; an automatic tube current control mechanism was applied to deliver a tube current of 300 mA for a body weight of 70-80 kg. CT images were then acquired with X-ray tube rotation time of 0.5 seconds. Treatment planning was performed using radiotherapy planning system Oncentra Brachy (Elekta, Sweden) version 4.6.0 with the acquired CT images. The first dwell position of the source was 1,500 mm from the source storage, and the second was 10 mm from the first dwell position. Irradiation was conducted according to the treatment plan using a microSelectron HDR v.3 (Elekta, Sweden). Air-kerma rate of the source was 19.712 mGy-m2/h at the time of the study. The source was dwelled for 20 seconds at each position.

The following imaging conditions were examined. In arrangement (1), the longitudinal axis of the applicator was parallel to the travel direction of the CT gantry, and the tip of the applicator was in the center of the gantry. CT scout images were acquired, while an 192Ir source microSelectron v.2c (Elekta, Sweden) was in the applicator under conditions detailed in Table 1. The effects of different imaging conditions and applicator geometry were examined on the accuracy of source detection. All tube voltages were set to 120 kVp. Arrangements (2) and (3) were used to examine whether the proposed method affects the source detection when the applicator was not in the center of CT gantry. In arrangement (4), an actual clinical situation was simulated; imaging of an applicator inserted into a patient at a fixed position is difficult in practice, hence it must be imaged on an arbitrary position. As presented in Figure 2, X-ray tube was oriented TOP when it was above the bed, BOT when it was below the bed, and LAT when it was on the right-hand side of the bed (when viewed from the applicator tip).

Fig. 2

CT scout image acquisition indicating imagingdirections. TOP, LAT, and BOT images were taken fromupper, right side, and bottom sides in the CT gantry, respectively

/f/fulltexts/JCB/51750/JCB-15-51750-g002_min.jpg
Table 1

CT scout imaging conditions and applicator geometries for four different arrangements

ArrangementWater phantom installation positionTube current
(1) StandardGantry center35 mA
(1-b)20 mA
(1-c)45 mA
(2)Couch lowered 130 mm from the standard35 mA
(3)50 mm laterally from (1)
in the direction away from
the tube during LAT imaging
(4)No guide near the center
of CT gantry

Source dwell position detection program

A program to detect the source dwell position from CT scout images was created using MATLAB 2019a (MathWorks, MA, United States); this program analyzed coordinates of the first and second dwell positions of the source in three dimensions by combining either TOP or BOT images, which were in XZ plane with LAT image, which was in YZ plane, as shown in Figure 3.

Fig. 3

Axial directions with respect to experimental phantom set-up. X and Z directions are vertical and parallel, respectively, to the direction of CT gantry movement direction. Y direction is from lower to upper side of the water phantom

/f/fulltexts/JCB/51750/JCB-15-51750-g003_min.jpg

Figure 4 demonstrates a representation of MATLAB program. An image of the source at the first dwell position was created by subtracting an image taken at the second dwell point from a loaded image taken at the first dwell point. Similarly, an image of the source at the second dwell position was created by subtracting an image taken at the first dwell point from an image taken at the second dwell position. Binarization was performed with a pixel value threshold of 90. This value was determined to be the threshold pixel value, at which the source and the rest of the range could be distinguished based on an analysis of a combination of TOP and LAT images from the standard arrangement (1). Detection was carried out in a 6 pixels squared range, starting from the upper left of the image when the signal value was 1, and concluding at the lower right edge when the range reached 0. In the detected range with a signal value of 1, pixels were weighted by the original pixel value. The weighted center of the range was then used as a coordinate of the source dwell position. The number of pixels in the image was 512 pixels squared, with 1 mm/ pixel. In addition, an 8-direction labeling process was applied, so that the source could be recognized even if it was not placed parallel to a specific axis. The detected source dwell coordinates were only compared with the treatment plan in the z-direction. The source dwell position of the treatment plan was obtained in digital imaging and communications in medicine (DICOM) format. In accordance with AAPM TG56 guidelines [20], the system was judged appropriate when the coordinates of the source detected by the program were within ± 1 mm of those detailed in the treatment plan.

Fig. 4

Overview of source detection program. This figure shows the flow of source position detection in the tandem tube. The first two photographs from different directions are loaded. Source position is parallel to the CT gantry running direction

/f/fulltexts/JCB/51750/JCB-15-51750-g004_min.jpg

Results

Effects of different imaging conditions

The source was visually confirmed to be inside the applicator under all conditions in CT scout images acquired in this study. The average of subtracted pixel values in the range taken to include the applicator tip, source, and surrounding pixels against standard arrangement (1) were ±0.42 for (1-b) and ±0.47 for (1-c). Figure 5 shows BOT images and pixel values for arrangements (1) and (2). Pixel values corresponded to the applicator tip and source area for different phantom placement positions. The image of the water phantom in Figure 5B was enlarged. The width of the water phantom lid fastener, indicated by a two-way arrow, increased by 127%, from 8.6 mm to 10.9 mm. The enlargement of the image in the x-direction was visually clear in BOT images. Close to the applicator tip, the positive range of pixel values was one pixel wider in the x-direction (Figure 5B), as indicated by the bold rectangle.

Fig. 5

BOT images and pixel values for standard arrangement (1) and (2). High pixel value areas are surrounded by bold lines. These areas indicate the source and applicator

/f/fulltexts/JCB/51750/JCB-15-51750-g005_min.jpg

Comparison of source dwell position coordinates obtained from detection and the treatment plan

The (x, z) coordinates were obtained from the analysis of TOP and BOT images, and the (y, z) coordinates were obtained from the analysis of LAT images using the in-house MATLAB program. Table 2 shows the difference between z-coordinates obtained from each image from the in-house program and the treatment plan. The mean and standard deviation (1 SD) of the difference were –0.47 ±0.25, –0.24 ±0.20, and –0.39 ±0.19 mm for TOP, LAT, and BOT images, respectively. Values exceeding this range were found in TOP images at the first dwell position and BOT images at the second dwell position for arrangement (1-c), and in LAT images at the second dwell position in arrangement (2). All values were within ±1 mm under all conditions.

Table 2

Difference between program output and treatment plan coordinates

Imaging
conditions
Difference from treatment plan in z-direction (mm)
Dwell pointTOP (XZ plane)LAT (YZ plane)BOT (XY plane)
(1)First dwell point–0.12–0.00–0.38
Second dwell point–0.27–0.07–0.30
(1-b)First dwell pointNot detectable–0.31–0.38
Second dwell pointNot detectable–0.11–0.32
(1-c)First dwell point–0.91–0.33–0.41
Second dwell point–0.49–0.28–0.83
(2)First dwell point–0.22–0.15–0.57
Second dwell point–0.64–0.73–0.38
(3)First dwell point–0.58–0.27–0.29
Second dwell point–0.71–0.16–0.44
(4)First dwell point–0.06–0.04–0.04
Second dwell point–0.36–0.43–0.39

The source coordinate was undetectable in the analysis of TOP and LAT images for arrangement (1-b). In addition, the analysis of TOP and LAT images for arrangement (1-c) output more coordinates than the number of actual dwell positions. In Table 2, the nearest value to the z-coordinate of the treatment plan is shown. According to periodic quality control at our institute, the source was confirmed to be within ±0.5 mm of the designation of irradiation equipment using a source dwell position confirmation tool (Elekta, Sweden).

Discussion

Differences in ct scout images due to different imaging conditions

Variations in the tube current did not result in differences in pixel values of CT scout images. Therefore, we can conclude that our program can detect the source regardless of imaging conditions, when the tube current is within the range of 20-45 mA. Previous studies have reported that the object to be imaged appears vertically elongated in images when it is not in the gantry center or extremely far from X-ray tube, and horizontally elongated when it is close because of the magnification ratio [21]. In the phantom position arrangement (2), the applicator was located far from the X-ray tube. In TOP image in arrangement (2), the image was enlarged in the z-coordinate of the applicator. However, no differences were observed in the position or number of pixels with high values, suggesting that the magnification in the z-direction was less than 1 mm. In arrangement (3), the applicator was positioned at the edge of both TOP and BOT images. The image was enlarged in the x-coordinate in BOT image, with one additional row of pixels with a positive value, indicating the presence of the applicator. Therefore, the expansion in the x-coordinate was greater than 1 mm. This enlargement was caused by a fan beam. Scout images are produced using a fan beam X-ray collimated in 0.6 mm/detector in the center of the gantry. In the z-direction, the X-ray width was 0.6 mm, which was very narrow; therefore, the impact for enlargement was very small. In the x-direction, 6 detectors that were lined up in an arc were applied. The enlargement occurred in the x-direction for TOP and BOT images, and in the y-direction for LAT images. Therefore, the positions in x- and y-directions were inaccurate. In addition, depending on the distance from the gantry center, objects being imaged could be magnified or demagnified. Therefore, in this study, only z-direction coordinates analyzed by the system were compared with those of the treatment plans.

Detection of source dwell position using in-house matlab program

The source dwell position was within ±1 mm of that defined in the treatment plan in all images, which was within the tolerance limit for dwell position accuracy recommended by AAPM TG56 [20]. However, even though they were within acceptable limits, the coordinates of the detected source dwell position tended to have a negative displacement with respect to the treatment plan, that is, closer to the connection between the applicator and irradiation equipment. This may be because of the structure of the source. The core of the source is sealed inside a stainless-steel capsule, and the center of the source core is 0.25 mm closer to the tip than the center of the capsule. Coordinates of the source dwell position in the planning system were defined as the center of the source core. In contrast, the proposed method that considers a range of high pixel values as the source, outputs the center of the source capsule as the coordinates of the source center. Therefore, the value obtained by our proposed method would be negative compared with that defined in the treatment plan. When investigating the effect of different imaging conditions, the analysis became impossible when the binarization threshold value was 90 for (1-b) and (1-c) arrangements. This might be because several points were mis-detected as sources. The source of the subtraction image produced for analysis consisted of a series of five pixels values exceeding the threshold of 90. The distance per pixel was 1.0 mm, indicating that a series of five pixels covers a larger area than the actual source. The program is designed not to recognize a single pixel or more than five pixels that exceed the threshold value as a source to prevent false detections, which may be the reason for the source not being detected by the program. As shown in Figure 6, another reason for misdetection was that the pixels with values exceeding 90 were next to each other in the area corresponding to the applicator in TOP images for arrangement (1-c), resulting in the recognition of multiple sources. Figure 6 shows a subtracted image for detecting the source at the second source position. In the image, there are two high pixel value areas, one is a part of the applicator wall around the first position and the other is the second source position. When the binarization threshold was 90, the system detected both areas as a source position. Some pixel values also exceeded 90 in TOP subtraction image in arrangement (1), as shown in Figure 7. However, these were not recognized as sources because the neighboring pixels did not exceed the value of 90.

Fig. 6

TOP subtraction image for detecting second source position in arrangement (1-c) and corresponding pixel values. Bold lines indicate applicator and the second dwell position. Dotted rectangle area indicates the first dwell position. There are some pixels with high positive pixel values. Along applicator wall around the first dwell position, two pixels over 90 exist next to each other

/f/fulltexts/JCB/51750/JCB-15-51750-g006_min.jpg
Fig. 7

TOP subtraction image for detecting second source position in arrangement (1) and corresponding pixel values. Only one pixel over 90 is present at around first dwell position

/f/fulltexts/JCB/51750/JCB-15-51750-g007_min.jpg

To properly analyze the undetectable CT scout image combinations, the threshold value was increased to 100 for TOP and LAT images for arrangements (1-b) and (1-c). With this change, the coordinates of source dwell position were detected. The analysis was successful when the threshold was increased to 100, because pixels with a value of over 100 only existed individually in the remaining portion of the applicator and were not identified as a source. In the effect of different imaging conditions section, we concluded that the imaging conditions do not result in a significant difference in the pixel values of the images; however, the images were undetectable for arrangements (1-b) and (1-c). Therefore, in the proposed system, small differences in pixel values affect source detection, and the results of this study are comparable to those reported in the existing literature. Watanabe et al. reported a difference between the simulated source position of the check cable and the source dwell position analyzed by a pinhole camera of 1.5 ±0.7 mm in longitudinal direction of the source [15]. In the report about an analysis of images acquired using the sources by an electronic portal imaging device for a quality control, the difference with an average of 0.26 mm and a maximum of 0.5 mm was shown between the observed source dwell position and that of the treatment plan [17]. However, these reports analyzed source dwell intervals. In contrast, in the current study, we analyzed the source dwell position in terms of absolute coordinates. The ICRP report of erroneous irradiation cases [5] reported some cases, in which the dwell position in the applicator was different from that in the treatment plan, because the length of the applicator was not entered accurately in the treatment plan. The system developed in this study may be able to detect these cases to be detected because it considers absolute coordinates. Studies on instances, which the analysis failed could lead to more accurate analysis that can be used in clinical practice.

Conclusions

Quality control is performed regularly at the facility where this study was conducted. Quality control using a proper tool has confirmed that the source can be stopped within ±1 mm of the target position. Comparison of coordinates of the source dwell position with the treatment plan using the in-house MATLAB program and CT scout images was possible with an accuracy of within ±1 mm. The source dwell position can be properly analyzed by acquiring images as close to the gantry center as possible to avoid applicator enlargement in the CT scout image. The system proposed in this study can be easily implemented using existing equipment for image-guided brachytherapy, and can analyze the absolute coordinates of the source dwell position. This can prevent inaccurate irradiation by analyzing whether HDR brachytherapy was performed properly. The system should be upgraded for more studies and for clinical use.

Disclosure

This research received a research grant from the Japan mHDR Research Fund.

References

1 

ICRU Report 38, Dose and Volume Specification for Reporting Intracavitary Therapy in Gynecology. International Commission on Radiation Units and Measurements, Bethesda 1985.

2 

Fredh A, Scherman JB, Fog LS et al. Patient QA systems for rotational radiation therapy: A comparative experimental study with intentional errors. Med Phys 2013; 40: 031716.

3 

Berbeco RI, Neicu T, Rietzel E et al. A technique for respiratory-gated radiotherapy treatment verification with an EPID in cine mode. Phys Med Biol 2005; 50: 3669-3679.

4 

Kubo HD, Glasgow GP, Pethel TD et al. High dose-rate brachytherapy treatment delivery: Report of the AAPM Radiation Therapy Committee Task Group No. 59. Med Phys 1998; 25: 375-403.

5 

Valentin J. Prevention of high-dose-rate brachytherapy accidents. ICRP Publication 97. International Commission on Radiological Protection, 2005.

6 

United States Nuclear Regulatory Commission Office of Federal and State Materials and Environmental Management Programs. NRC Information Notice 2012–08, 2012.

7 

Duan J, Macey DJ, Pareek PN et al. Real-time monitoring and verification of in vivo high dose rate brachytherapy using a pinhole camera. Med Phys 2001; 28: 167-173.

8 

Nakano T, Suchowerska N, Bilek M et al. High dose-rate brachytherapy source localization: Positional resolution using a diamond detector. Phys Med Biol 2003; 48: 2133-2146.

9 

Song H, Bowsher J, Das S et al. Tracking brachytherapy sources using emission imaging with one flat panel detector. Med Phys 2009; 36: 1109-1111.

10 

Kojima H, Hanada T, Katsuta S et al. New method for obtaining position and time structure of source in HDR remote afterloading brachytherapy unit utilizing light emission from scintillator. J Appl Clin Med Phys 2009; 10: 86-95.

11 

Smith RL, Taylor ML, McDermott LN et al. Source position verification and dosimetry in HDR brachytherapy using an EPID. Med Phys 2013; 40: 111706-1170612.

12 

Miyahara Y, Kitagaki H, Nishimura T et al. Usefulness of direct-conversion flat-panel detector system as a quality assurance tool for high-dose-rate 192Ir source. J Appl Clin Med Phys 2015; 16: 121-129.

13 

Nose T, Chatani M, Otani Y et al. Real-time verification of a high-dose-rate iridium 192 source position using a modified C-arm fluoroscope. Int J Radiat Oncol Biol Phys 2017; 97: 858-865.

14 

Poder J, Cutajar D, Guatelli S et al. HDR brachytherapy in vivo source position verification using a 2D diode array: A Monte Carlo study. J Appl Clin Med Phys 2018; 19: 163-172.

15 

Watanabe Y, Muraishi H, Takei H et al. Automated source tracking with a pinhole imaging system during high-dose-rate brachytherapy treatment. Phys Med Biol 2018; 63: 145002.

16 

Belley MD, Craciunescu O, Chang Z et al. Real-time dose-rate monitoring with gynecologic brachytherapy: Results of an initial clinical trial. Brachytherapy 2018; 17: 1023-1029.

17 

Nose T, Masui K, Takenaka T et al. An easy and novel method for safer brachytherapy: Real-time fluoroscopic verification of high-dose-rate 192Ir source position using a flat-panel detector. J Radiat Res 2019; 60: 412-415.

18 

Yogo K, Noguchi Y, Okudaira K et al. Source position measurement by Cherenkov emission imaging from applicators for high-dose-rate brachytherapy. Med Phys 2021; 48: 488-499.

19 

Mutic S, Palta JR, Butker EK et al. Quality assurance for computed-tomography simulators and the computed-tomography-simulation process: Report of AAPM Radiation Therapy Committee Task Group No. 66. Med Phys 2003; 30: 2762-2792.

20 

Ravinder N, Lowell LA, Jerome AM et al. Code of practice for brachytherapy physics: Report of the AAPM Radiation Therapy Committee Task Group No. 56. Med Phys 1997; 24: 1557-1598.

21 

Sukupova L, Vedlich D, Jiru F. Consequences of the patient’s mis-centering on the radiation dose and image quality in CT imaging-phantom and clinical study. Univers J Med Sci 2016; 4: 102-107.

Copyright: © 2023 Termedia Sp. z o. o. This is an Open Access article distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 International (CC BY-NC-SA 4.0) License (http://creativecommons.org/licenses/by-nc-sa/4.0/), allowing third parties to copy and redistribute the material in any medium or format and to remix, transform, and build upon the material, provided the original work is properly cited and states its license.
 
Quick links
© 2024 Termedia Sp. z o.o.
Developed by Bentus.