A new laboratory device with mathematically based positioning of a frozen tissue block facilitating precise sectioning of large specimens

Andrzej Bieniek
,
Łukasz Matusiak
,
Zdzisław Woźniak
,
Maciej Lichtenstein
,
Janusz Szymkowski
,
Maria Kozioł
,
Anna Salwa
,
Jacek C. Szepietowski

Pol J Pathol 2016; 67 (2): 151-155
Online publish date: 2016/08/02
Article file
- A new.pdf  [0.19 MB]
Get citation
 
PlumX metrics:
 

Introduction

Mohs micrographic surgery (MMS) constitutes the method of surgical treatment of skin cancers assuring the highest cure rates with possible preservation of healthy tissue. It aims at thorough, personalized eradication of skin cancers by mean of staged excision of tissues surrounding the tumor with complete (100%) histopathological examination of their margins. Tissue flattening and mounting in a parallel alignment to the surface of the object holder and the microtome knife section plane aims at cutting the most superficial sections of a specimen block, thus enabling the complete histopathological examination of excision margins of skin cancers. Numerous methods and devices have been designed to facilitate these key steps of tissue processing in MMS. Some of these methods are: manual (direct) technique, microscope slide technique, Bard-Parker scalpel, freezing bar technique, plastic “Cryomolds”, “Davidson Cryocup” metal molds, Carter’s device, Koehn’s modification of Carter’s device, “Miami Special” clamps, Gormley’s device, Motley & Holt’s device, Marini’s “CryoSystem”, Franks’ device, and “CryoHist” [1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14]. Some inaccuracies of tissue processing are unavoidable and depend on the unique features of a particular device or technique. The suboptimal specimen orientation is subsequently corrected by the manual adjustment of the object holder in a cryostat [2, 3]. However, it might not only prove difficult and time-consuming but also may be associated with errors: cutting the incomplete sections or cutting the section at considerable depth. This may result in overlooking cancer or unnecessary removal of healthy tissue. Small organs, e.g. eyelids, are particularly prone to such inaccuracies.

Objective

The aim of the study was to design a device which enables exact mechanical adjustment of a specimen block surface of diameter up to 50 mm, to the current position of a microtome knife, as well as flattening, freezing, embedding and mounting of the tissue in MMS.

Material and methods

We designed a prototype device for tissue processing in MMS. Its freezing system consists of a liquid nitrogen container and an outlet hose with a footoperated magnetic valve. The main body of the device consists of a base (Fig. 1A), an arm supporting a stage (Fig. 1B) moved by an adjustment knob (Fig. 1C), with clamps for microscopic slides (of standard size: 25 × 70 mm or larger: 50 × 70 mm) (Fig. 1D), a mirror (Fig. 1E), a transparent shield (Fig. 1F) and a socket for the specimen holder (Fig. 1G) placed above the gas outlet (Fig. 1H). The socket is surrounded by three threaded mandrels (Fig. 2); the two frontal ones (left and right) are regulated by knurled knobs (Fig. 3). Iron-core specimen holders of 50 mm in diameter are pulled by a magnet fixed in the gas outlet, which provides the necessary stabilization. The mechanical, mathematically based adjustment of a plane of tissue block to the actual position of the knife is carried out in three steps:
Step 1: Preparation of the “sample block” from the frozen OTC medium (without tissue). A specimen holder is inserted into the device socket with the mark indicating 6 o’clock. After the freezing process is released, OTC medium is poured on its surface and pressed with a glass slide mounted on a stage. When the slide is removed, a tissueless sample block is ready.
Step 2: Sectioning the sample block in a cryostat microtome. The sample block is secured in a cryostat socket, with its mark indicating 6 o’clock. Sections of 10 m are cut. Three parameters should be taken into consideration: I) Position of point on the circumference of the block where the sectioning begins (point A). It should be indicated by the clock – in order to determine it accurately, the special ruler in the form of a dial may be applied. II) Depth of first complete section (in micrometers) – a distance from the block surface which equals the number of the cryostat crank turns multiplied by 10. III) Diameter of the block (in millimeters) between points A and B (point B is situated opposite to A, and constitutes the finally cut fragment of the full face section).
Step 3: Adjusting the position of a specimen holder by means of the rotating mandrels. The results of the described parameters are entered into a computer program (Block Adjustment Calculator) which in a fraction of a second provides the necessary shift of mandrels. The adjustment process should be repeated each time when the cryostat chuck position or the settings of the device were presumably altered. To be sure we repeat it every morning before work.
After completing the adjustment, the tissue processing is carried out in the following way: A glass slide is placed on the stage. Tissue (with relaxing cuts if needed) is spread on the slide, flattened, pressed and frozen under visual control (the rear part of the specimen is visible in the mirror) with individually regulated speed. It is also recommended to check if its bottom surface adheres strictly to the slide (it is visible through the glass). After complete freezing of tissue, its edges may be covered with a wreath of medium, which prevents their further detachment from the slide. Then the specimen holder is secured in the device socket with the mark at 6 o’clock and covered with the OTC medium. The glass slide is rotated tissue down and pressed to the OCT medium. When the specimen block is completely frozen, the glass is removed and the object disc is inserted into the cryostat chuck, with the mark indicating 6 o’clock. Then the tissue block is ready for precise sectioning without any additional manual adjustment. The only limitation of the size of sections is the diameter of the object holder, which is actually 50 mm.
The device was used in 212 MMS operations of skin cancer, by which 513 histological slides from tumor margins were prepared. The tissue specimens varied in dimensions and composition, containing skin, fat, muscle fascia or cartilage. The following parameters were measured: the depth of the first complete sections and the largest diameter of specimens.

Results

The full face sections were obtained at average depth of 81.60 m (min. 20, max. 180, SD = 29.15), whereas the average diameter of sections was 18.11 mm (min. 4 mm, max. 42 mm, SD = 9.10).

Discussion

We have performed the “fresh tissue” Mohs micrographic surgery since 1999. Initially, we used the “direct method” of tissue processing using a heat extractor and cryostat equipped with an articulated chuck. However, this method in our practice proved difficult to control and was burdened with considerable imperfections. Great improvements followed the implementation of a new device – Cryosystem [12]. However, major drawbacks continued to occur with the use of the cryostat together with another laboratory, which forced us to perform manual adjustment of the block surface in relation to the plane of the blade. In this situation, we started research on the mechanical alignment of the tissue block, which after many attempts led to an efficient device. Its usage proved to be a considerable relief, helping us to markedly facilitate the procedure and increase its accuracy. Later our laboratory was equipped with a cryostat with a fixed chuck, which is intended exclusively for MMS. We noticed that such a mechanism is more stable than an articulated chuck, which positively affected the quality of sectioning. However, its factory settings were not set in an ideal position in relation to the plane of the microtome knife. Hence we continued to use the mechanical adjustment with our device.
We have used the device in 212 MMS operations, in which two kinds of cryostats – with a fixed or articulated chuck – were employed. It proved that complete sections were obtained close to the surface of the tissue block, at an average depth of 81.60 µm. What is important, such results were obtained despite the relatively large size of specimens. (The average diameter was as high as 18.11 mm. The diameter of histological specimens was smaller than the tissue in vivo by 20-30%, because of their shrinkage.)
It is thought that the applied method is not as crucial in MMS processing as the experience and scrupulousness of a histotechnician [15]. However, a majority of tissue processing methods do not provide the highest expected precision, even in the most experienced hands. The best accuracy is delivered by the Franks’ device [13] and the CryoHist machine [6]. Both of them facilitate obtaining full-face sections at an average depth not exceeding 100 µm, on condition only that the position of the cryostat chuck is constantly fixed. Our device assures similar precision, but is applicable also in cryostats in which the position of the articulated chuck is being changed, for instance by other users. Nevertheless, the degree of precision of most methods is far from that and equals several hundred micrometers. For example, the commonly used direct method may cause inaccuracies of even 2 mm [9]. These drawbacks are well known to specialists – inaccuracies in histological processing are not only troublesome but also may result in producing histological slides which may be ineligible for diagnostic purposes [16, 17]. Our measurements and observations proved that our own designed prototype device is comparable to the most precise methods described so far, and is far more exact than most of the commonly used procedures.
Moreover, the histological processing in MMS is in the majority of its present forms complicated. In consequence, it can be performed only by highly specialized personnel trained for a long time in departments which have a cryostat for their exclusive use. Our device allows one to obtain high quality sections by performing simple, clearly defined actions. Brief training is sufficient – it could be ascertained that after 1 hour technicians were able to operate the device effortlessly with good results. This ease causes that tissue may be processed in a short time; the adjustment usually takes less than 4 minutes, whereas mounting and sectioning takes around 2 minutes.
Especially useful in large tumors is the possibility of creating large histological specimens, with the use of oversized object discs and slides enabled by the device (Fig. 4). This allows one to avoid failures associated with tissue division, such as pushing malignant structures toward the undersurface plane (false positive specimens), as well as orientation errors [18, 19, 20, 21]. This feature may be helpful not only in the treatment of skin cancer using Mohs surgery, but also during the surgical treatment of neoplasms located within other organs, where “whole mount histological specimens” are in use, namely head and neck, breast, prostate, brain, etc.
In summary, the device may prove especially useful in Mohs centers using the direct technique or other techniques in tissue processing, requiring manual adjustment of the cryostat chuck position. It can also be useful in laboratories equipped with a cryostat with a fixed position chuck, aiming at correction of its minimal inaccuracies. The device may also enhance the performance of those Mohs surgery departments which share a cryostat with other units. Especially, it could facilitate MMS in traditional hospital histopathology laboratories without acquiring a separate cryostat.

The authors declare no conflict of interest.
Works on the laboratory device were financed partially from a grant of the Polish Ministry of Science and Higher Education, no. 701/2008, entitled “Computer data analysis in Mohs micrographic surgery – method of skin cancer treatment”.
Andrzej Bieniek and Janusz Szymkowski are the authors of a patent on the device, submitted to the Polish Patent Office under number P398733.

References

1. Dogan MM, Snow SN, Lo J. Rapid skin edge elevation using the OCT compound droplet technique to obtain horizontal microsections in Mohs micrographic surgery. J Dermatol Surg Oncol 1991; 17: 857-860.
2. Selak LS. The Mohs Laboratory. In: Mikhail G (ed.). Mohs micrographic surgery. Saunders Phildelphia 1991; 299-308.
3. Yob EH. Cryostat and cryostat technique: Obtaining Quality Mohs Slides. In: Gross KG, Steinmann HK, Rapini RP (eds.). Mohs Surgery: Fundamentals and Techniques. Mosby, St Louis, Baltimore Boston 1998.
4. Miller LJ, Argenyi ZB, Whitaker DC. The preparation of frozen sections for Mohs micrographic surgery. A review of current metodology. J Dematol Surg Oncol 1993; 19: 1023-1029.
5. Silapunt S, Peterson R, Alcalay J, Goldberg LH. Mohs tissue mapping and processing: a survey study. Dermatol Surg 2003; 29: 1109-1112.
6. Hanke CW, Leonard AL, Reed AJ. Rapid preparation of high quality frozen sections using a membrane and vacuum system embedding machine. Dermatol Surg 2008; 34: 20-25.
7. Hanke CW, Lee MW. Cryostat use and tissue processing in Mohs micrographic surgery. J Dermatol Surg Oncol 1989; 15: 29-32.
8. Carter VH. A new method for preparing tissue blocks for cryostat sectioning. J Dermatol Surg Oncol 1985; 11: 687-689.
9. Gormley DE. Evaluation of a method for controlled tissue embedding for histologic evaluation of tumor margins. Am J Dermatopathol 1987; 9: 308-315.
10. Koehn GG. A new modification for preparing tissue blocks for cryostat sectioning. J Dermatol Surg Oncol 1992; 18: 485-486.
11. Motley RJ, Holt PJ. A simple device for optimal tissue preparation for Mohs micrographic surgery. Br J Dermatol 1992; 126: 57-59.
12. Marini L. Cryosystem: a new, complete device for tissue processing in micrographic surgery. J Eur Acad Dermat Venereol 1995; Supl. 1: 117.
13. Franks JW. A precision machine for mounting tissues for Mohs micrographic surgery. Dermatol Surg 1998; 24: 989-993.
14. Peters S. The Art of Embedding Tissue for Frozen Section, in: Stephen R. Peters (ed): A Practical Guide to Frozen Section Technique. Springer-Science+ Business Media, Pathology Innovations LLC 2010 [http://www.leicabiosystems.com/pathologyleaders/the-art-of-embedding-tissue-for-frozen-section/].
15. Tulli A. Mohs’ micrographic surgery. In: Chu AC, Edelson RL (eds.). Malignant Tumors of the Skin. Oxford University Press, London, UK 1999; 381-395.
16. Rapini RP. Pitfalls of Mohs micrographic surgery. J Am Acad Dermatol 1990; 22: 681-686.
17. Soo Hoo W, Ruebner B, Vogt P, Wiese D. Orientation of small, flat, frozen-section specimens. Am J Surg Pathol 1988; 12: 573-574.
18. Randle HW, Zitelli J, Brodland DG, Roenigk RK. Histologic preparation for Mohs micrographic surgery. The single section method. J Dermatol Surg Oncol 1993; 19: 522-524.
19. Sukal SA, Tudisco M, Strippoli B, Nehal KS. Innovative laboratory techniques to facilitate processing of large Mohs cases. Dermatol Surg 2005; 31: 763-765.
20. Gloster HM. Surgical pearl: Large single sections in Mohs micrographic surgery. J Am Acad Dermatol 2003; 49: 506-508.
21. Ellis JI, Khrom T, Wong A, et al. Mohs math-where the error hides. BMC Dermatol 2006; 6: 10.

Address for correspondence

Andrzej Bieniek, Assoc. Prof., MD, PhD
Department of Dermatology, Venereology and Allergology
Wroclaw Medical University
Chałubińskiego 1
50-368 Wrocław, Poland
tel. +48 71 784 22 86
fax +48 71 327 09 42
e-mail: andrzej.bieniek@am.wroc.pl
Copyright: © 2016 Polish Association of Pathologists and the Polish Branch of the International Academy of Pathology This is an Open Access article distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 International (CC BY-NC-SA 4.0) License (http://creativecommons.org/licenses/by-nc-sa/4.0/), allowing third parties to copy and redistribute the material in any medium or format and to remix, transform, and build upon the material, provided the original work is properly cited and states its license.
Quick links
© 2024 Termedia Sp. z o.o.
Developed by Bentus.