eISSN: 2081-2841
ISSN: 1689-832X
Journal of Contemporary Brachytherapy
Current Issue Archive Supplements Articles in Press Journal Information Aims and Scope Editorial Office Editorial Board Register as Author Register as Reviewer Instructions for Authors Abstracting and indexing Subscription Advertising Information Links
Editorial System
Submit your Manuscript
SCImago Journal & Country Rank

2/2019
vol. 11
 
Share:
Share:
abstract:
Original paper

Development and assessment of 3D-printed individual applicators in gynecological MRI-guided brachytherapy

Helena Barbara Zobec Logar
1, 2
,
Robert Hudej
1
,
Barbara Šegedin
1, 2

1.
Department of Radiation Oncology, Institute of Oncology Ljubljana, Ljubljana, Slovenia
2.
Medical Faculty, University of Ljubljana, Ljubljana, Slovenia
J Contemp Brachytherapy 2019; 11, 2: 128–136
Online publish date: 2019/04/26
View full text Get citation
 
Purpose
To evaluate the clinical use of 3D printing technology for the modelling of individual applicators for advanced gynecological tumors in magnetic resonance imaging (MRI)-based brachytherapy (BT).

Material and methods
We tested individually designed 3D-printed applicators in nine patients with advanced gynecological cancer. Before BT was performed, all patients were treated with external beam radiotherapy (EBRT). The most common indication for individualized BT was advanced gynecological tumors where the use of standard BT applicators was not feasible. Other indications were suboptimal dose-volume histogram (DVH) parameters for high-risk clinical target volume (CTV-THR) at the first BT (V100 ≤ 90% of CTV-THR volume and D98 ≤ 80%, D90 ≤ 100%, and D100 ≤ 60% of dose aim). The EQD2 dose aim to the target volume D90 CTV-THR per one BT fraction was 20 Gy for cervical or recurrent endometrial cancer and 16 Gy for vaginal cancer patient. The first BT with the standard applicator in situ was used as the virtual plan for designing a 3D-printed applicator. The next BT was performed with a 3D-printed applicator in situ. The primary endpoint was to improve CTV-THR DVH parameters without exceeding the dose to the organs at risk (OARs).

Results
All DVH parameters for CTV-THR were significantly higher with the use of an individually designed applicator. Mean D90 CTV-THR improved from 14.1 ±5.4 Gy to 22.0 ±2.5 Gy and from 7.1 Gy to 16.2 Gy for cervical/recurrent endometrial and vaginal cancer, respectively (p < 0.001). The mean D2cm3 bladder, rectum, sigmoid, and bowel dose was within institutional dose constraints, and increased from 13.0 ±1.5 Gy to 13.6 ±1.5 Gy (p = 0.045), 10.8 ±1.2 Gy to 11.7 ±1.3 Gy (p = 0.004), 8.9 ±3.2 Gy to 10.3 ±3.3 Gy (p = 0.008), and 8.7 ±3.8 Gy to 9.2 ±3.1 Gy (p = 0.2).

Conclusions
With the use of individual 3D-printed applicators, all DVH parameters for CTV-THR significantly improved without compromising the dose constraints for the OARs.

keywords:

image-guided brachytherapy, 3D printing, gynecological cancer

 
Quick links
© 2024 Termedia Sp. z o.o.
Developed by Bentus.