Journal of Contemporary Brachytherapy
eISSN: 2081-2841
ISSN: 1689-832X
Journal of Contemporary Brachytherapy
Current Issue Archive Supplements Articles in Press Journal Information Aims and Scope Editorial Office Editorial Board Register as Author Register as Reviewer Instructions for Authors Abstracting and indexing Subscription Advertising Information Links
Editorial System
Submit your Manuscript
SCImago Journal & Country Rank

6/2025
vol. 17
 
Share:
Share:
abstract:
Original paper

Dosimetric comparison between manual and inverse optimization methods in high-dose-rate intracavitary/ interstitial cervical cancer brachytherapy for two treatment planning systems

Lisa Mahe
1
,
Leonel Varela Cagetti
2
,
Juliette Myxa
1
,
Laurence Gonzague-Casabianca
2
,
Pierre Fau
1
,
Hugues Mailleux
1
,
Agnès Tallet
2
,
Marjorie Ferre
1

  1. Department of Medical Physics, Institut Paoli-Calmettes, Marseille, France
  2. Department of Radiation Oncology, Institut Paoli-Calmettes, Marseille, France
J Contemp Brachytherapy 2025; 17, 6: 400–406
Online publish date: 2025/12/31
View full text Get citation
 
Purpose
To compare manual treatment planning with two inverse planning methods from two different treatment planning systems for high-dose-rate cervical cancer brachytherapy.

Material and methods
Thirty-one consecutive patients were included. Three treatment plans were created: manual, hybrid inverse treatment planning (HIPO), and RayStation Brachy (RSB), using Oncentra TPS for manual and HIPO, and RayStation TPS for RSB. Dose volume parameters, dwell time in intrauterine tandem, ovoids, and interstitial needles as well as Paddick, COIN, and dwell time inhomogeneity index, were evaluated.

Results
Compared with manual planning, HIPO plans resulted in a higher dose to targets while reducing the dose to organs at risk. It reduced the total treatment time and increased the loading in interstitial needles. RSB plans resulted in a similar dose to targets and organs at risk and a similar total treatment time, but a better homogeneity of dose distribution. All dosimetric indices were increased with inverse optimization.

Conclusions
Both inverse planning methods are able to produce dosimetrically acceptable plans, with an increase of target coverage and better homogeneity. RSB provides higher doses and better homogeneity than HIPO.

keywords:

brachytherapy, locally advanced cervical cancer, inverse optimization, HIPO, RayStation Brachy, high-dose-rate

 