Journal of Contemporary Brachytherapy

Abstract

4/2024 vol. 16
Original paper

Dosimetric evaluation and Monte Carlo simulation of a new proposed surface brachytherapy mould

  1. Department of Medical Physics, School of Medicine, Iran University of Medicine Sciences, Tehran, Iran
  2. Department of Radiation Oncology, Shohada Tajrish Educational Hospital, Medical School, Shahid Beheshti University of Medical Sciences, Tehran, Iran
  3. Cancer Institute, Tehran University of Medical Sciences, Tehran, Iran
  4. Medical Physics Department, Surrey University, Guildford, UK
  5. Department of Radiation Oncology, University of Utah, Salt Lake City, Utah, USA
J Contemp Brachytherapy 2024; 16, 4: 279-288
Online publish date: 2024/09/16
View full text

Purpose:

The aim of this study was to develop a new in-house low-cost surface mould, and to evaluate its performance and dosimetric properties for high-dose-rate (HDR) cobalt-60 (60Co) brachytherapy.

Material and methods:

A water-equivalent surface mould was developed using medical silicone. Mould performance and dosimetry characteristics were evaluated with Monte Carlo N-particle (MCNP2.6) simulation, Gafchromic™ EBT3 film measurements, and treatment planning system (TPS) output. Three sample moulds with different thicknesses (i.e., 0.5 cm, 1 cm, and 1.5 cm) were constructed, and a phantom study was performed. Treatment plans prescribing 3 Gy to 0.5 cm under pseudo-skin were designed, and film dosimetry was completed. TPS dose distributions were compared using Monte Carlo (MC) simulation and film dosimetry.

Results:

Good consistency was observed between TPS results and film dosimetry at the prescribed depth of 0.5 cm, with mean differences of 0.70%, 0.40%, and 0.19% for mould thicknesses of 0.5 cm, 1 cm, and 1.5 cm, respectively. However, higher discrepancies were found at the phantom surface with 1.00%, 0.80%, and 0.56% dose differences for the considered mould thicknesses, respectively. These increased differences could be due to a higher dose gradient at the phantom surface, and a greater impact of uncertainties on the obtained results in this part. Moreover, mean differences between the results obtained from MC simulations and output of TPS at the prescribed depth of 0.5 cm were 0.73%, 0.60%, and 0.08% for mold thicknesses of 0.5 cm, 1 cm, and 1.5 cm, respectively. Higher variations were observed between TPS and MC at the phantom surface with 1.30%, 0.70%, and 0.13% dose differences for the considered mould thicknesses, respectively.

Conclusions:

The developed surface mould demonstrated water equivalence at 60Co energies, and was consistent with TPS calculations at routine treatment depths. Its effectiveness in non-melanoma skin cancer (NMSC) lesion treatment is highlighted. However, due to mould attenuation, TG-43-based TPS overestimated the dose delivered using this mould, especially at pseudo-skin surface, emphasizing the necessity for a model-based TPS algorithm.