eISSN: 1897-4295
ISSN: 1734-9338
Advances in Interventional Cardiology/Postępy w Kardiologii Interwencyjnej
Current issue Archive Manuscripts accepted About the journal Editorial board Abstracting and indexing Subscription Contact Instructions for authors Publication charge Ethical standards and procedures
Editorial System
Submit your Manuscript
SCImago Journal & Country Rank
2/2022
vol. 18
 
Share:
Share:
abstract:
Image in intervention

A pacemaker lead indenting the wall of the left anterior descending artery

Łukasz M. Gabryel
1
,
Klaudyna Bugla
1
,
Jacek Jafra
2
,
Jacek Gabryel
2

1.
Department of Cardiology and Internal Medicine, School of Medicine, Collegium Medicum, University of Warmia and Mazury, Olsztyn, Poland
2.
Department of Cardiology, Voivodeship Specialist Hospital, Częstochowa, Poland
Adv Interv Cardiol 2022; 18, 2 (68): 178–179
Online publish date: 2022/08/19
View full text Get citation
 
An 81-year-old female with coronary disease, arterial hypertension, paroxysmal atrial fibrillation, and DDDR implantation 6 years ago due to sick sinus syndrome (SSS) was admitted to the cardiology ward with symptoms of stable angina pectoris to have invasive diagnostics performed. Selective coronarography revealed calcific ostial stenosis of the right coronary artery (RCA) and proximal stenosis of the left anterior descending artery (LAD). As well as the above-mentioned findings, it exposed the tip of an RV electrode with healing tissue around its spiral coil, which was indenting into the lumen of the middle part of the LAD causing stenosis and potential blood flow restriction (Figures 1 A–C). Further assessment of the LAD and the RCA stenoses using intravascular ultrasound (IVUS) and subsequent PCI of any lesion that would be considered significant were scheduled for the following month. During the second coronarography the exact mutual RV lead and the LAD’s position was determined by IVUS, which also showed no signs of damage to the artery wall (Figure 1 D). Considering the morphology of the LAD, the localization of indention in the distal part, the opinion of a cardiac surgeon, and our experience, we decided that the patient would not benefit from electrode removal or any other invasive treatment of the indented area. Therefore, we withdrew from invasive diagnostics and implemented conservative therapy with ambulatory observation. The stenosis of the ostial RCA was assessed in IVUS as nonsignificant (MLA = 6.5 mm2). However, the proximal, significant stenosis of LAD (MLA = 2.7 mm2) was treated with shockwave intravascular lithotripsy and DES stenting.
Cardiovascular implantable electronic devices cause various complications, and 0.2% of such procedures are complicated by cardiac tamponade, which is considered a rare incident [1]. In comparison to the above, damage of a coronary artery, with only a few cases reported [2], is an extremely rare complication. Narrowing it down to LAD injuries, there have been 4 cases reported [2, 3]. Nonetheless, many of the implantable leads remain in a very close relationship to the coronary arteries. The mean distance between them is 4.7 mm for the antero-septal junction fixation presented in this case [4]. Inferring from the study above and the analysed case, the optimal lead implantation technique for an RV electrode would be fixation on the true RV septum, and hence avoidance of antero-septal junction...


View full text...
Quick links
© 2024 Termedia Sp. z o.o.
Developed by Bentus.