Short communication

Percutaneous closure of perimembranous and postsurgical ventricular septal defects with Amplatzer Duct Occluder II Additional Sizes in paediatric patients – case series

Mateusz T. Knop
,
Linda Litwin
,
Małgorzata Szkutnik
,
Jacek Białkowski
,
Michal Galeczka
,
Roland Fiszer

Adv Interv Cardiol 2018; 14, 4 (54): 429–432
Online publish date: 2018/12/11
Article file
- Percutaneous closure.pdf  [0.08 MB]
Get citation
 
 

Introduction

Perimembranous ventricular septal defect (pmVSD), one of the most frequent congenital heart defects, requires surgical (gold standard) or transcatheter closure in a considerable number of children. In general, the results of surgical correction are excellent; however, in 2.5% of patients residual postsurgical VSDs (psVSD) are reported [1]. There have been numerous attempts to close pmVSD with various devices, with different efficacy and safety rates [2]. The percutaneous closure of psVSD, sometimes of atypical anatomy, was also reported in the literature [3]. The transcatheter treatment of pmVSD remains a controversial problem. The main issues are the risk of embolization, early and late conduction disturbances, damage to the tricuspid valve apparatus or aortic insufficiency [4]. Numerous devices have been proposed for percutaneous pmVSD closure, including the off-label use of Amplatzer muscular VSD occluders, Amplatzer duct occluders type I and II (ADO I, II) or ADOII Additional Sizes (ADO II AS) [2, 5–7]. To the best of our knowledge, there have been only two previous publications from this and the previous year regarding this application of ADO II AS in children – 1 case report [6] and 1 presentation of a series of 4 cases [7].

Aim

To present our preliminary experience with ADO II AS used for the transcatheter closure of 4 pmVSDs and 2 psVSDs.

Material and methods

All patients were qualified for percutaneous VSD closure by a team of paediatric cardiologists and cardiac surgeons based on clinical symptoms of increased pulmonary flow (chest X-ray) and/or echocardiographic signs of LV volume overload. All patients had a Qp/Qs ratio > 1.5. Written informed consent was obtained from the parents prior to the procedures.

Device

ADO II AS (St. Jude Medical, Inc., USA) is a device originally designed for arterial duct closure. Briefly, it is a symmetrical, self-expanding, single mesh layer Nitinol occluder. There are three different waist-disc diameters available (3 mm – 4 mm, 4 mm – 5.25 mm, 5 mm – 6.5 mm), each available with three different waist lengths (2, 4 and 6 mm). A dedicated Amplatzer TorqVue LP 4 French catheter is recommended for the deployment procedure. ADO II AS and its dedicated delivery catheter are characterized by soft construction – this feature is of special importance due to procedural manoeuvres carried out in the proximity of the conduction system and tricuspid valve apparatus. The device should be 1–2 mm bigger than the VSD diameter (as for patent ductus arteriosus cases).

Procedure

All procedures were carried out under general anaesthesia with elective intubation, fluoroscopy and direct transoesophageal echocardiography. Both venous and arterial femoral (4 Fr) access was obtained. Standard diagnostic catheterization was performed in all patients. The VSD was crossed from the LV in all cases with direct implantation of the ADO II AS from the LV in 3 children and from the RV – after arterio-venous loop creation – in another 3 patients. A dedicated Amplatzer 4 Fr delivery catheter was used in all cases.

Follow-up

All patients underwent Holter ECG monitoring and transthoracic echocardiography (TTE) on days 1 and 3 after the procedure. Antibiotic prophylaxis of endocarditis prophylaxis was continued for 6 months and extended if warranted by the presence of a residual shunt. Physical examination, ECG and TTE were performed 3, 6 and 12 months after VSD closure and yearly thereafter. The median follow-up was 10.5 months (range: 1–21 months).

Results

Six patients (2 males) were included in our analysis. Some clinical and procedural data of the treated patients are presented in Table I. The median age was 2.5 years (range: 1.3–8.8 years) and the median weight was 12.8 kg (range: 8.7–30 kg). Four children had an isolated pmVSD with a small aneurysm, the defect size ranging from 2.5 to 3.2 mm. Two patients with residual VSD previously underwent surgical correction of tetralogy of Fallot at the ages of 5 and 7 months and had a residual VSD of 3.2 mm (outflow type, patient 3) and 2.5 mm (Gerbode type, patient 4), respectively (Figure 1). In the first psVSD case an unsuccessful attempt to close the outflow VSD with the Amplatzer Duct Occluder II (a precursor of ADO II AS) was made 4 months before the final successful closure with ADO II AS. The pulmonary pressure was within normal limits in all patients.
There were no major adverse events in the periprocedural period, but we observed two transient moderate events. Soon after one of the procedures, we noted conduction disturbances with the progression of a partial right bundle branch block (RBBB) to complete RBBB with a new-onset left posterior fascicular block that resolved after a short course of steroid therapy (patient 1). Another patient with pmVSD (patient 6) had short-lasting, asymptomatic ventricular tachycardia (5 QRS complexes, 120/min) on the second day after the procedure. After device implantation, we did not observe any progression of tricuspid insufficiency or more than trivial aortic insufficiency. Two patients with psVSD had an insignificant residual shunt after the procedure that remained stable during further follow-up. No persistent or late-onset conduction abnormalities were observed.

Discussion

We present our initial data on the efficacy and safety of ADO II AS application in selected perimembranous and postsurgical VSDs. ADO II AS has several advantages – it has soft construction and its dedicated delivery system is delicate as well as requiring only a 4 Fr sheath. This increases the safety of manoeuvres inside the heart. Moreover, it is important to stress that ADO II AS can be easily implanted from the left side through the arterial access (without the creation of an arterio-venous loop) as it has a symmetrical construction and a relatively small size of retention discs. This was the case in 3 out of our 6 patients presented here and generally simplified the procedure (patients 4 and 5 required only approximately 9 min of fluoroscopy). The construction of ADO II AS ensures safe implantation despite the proximity of tricuspid and aortic valves and is believed not to interfere with the bundle of His. In the late follow-up, we have not observed any rhythm and conduction disturbances, nor aortic or tricuspid regurgitation in any of our cases. ADO II AS was originally designed for the percutaneous closure of patent ductus arteriosus (PDA) in small infants. This implant is particularly useful for the closure of connections less than 4 mm in diameter. We have also published our experience with ADO II AS for PDA closure in adolescents – in our centre this device substitutes coils for such purposes [8]. Champaneri et al. [6] have recently described the transcatheter closure of pmVSD in a 1.8 kg infant using ADO II AS (the age was not specified – probably in the second month of life). During the same procedure, residual postsurgical PDA was closed with another ADO II AS [6]. Narin et al. [7] presented their experience with percutaneous VSD closure in children under 1 year of age. There were 12 patients – ADOII was applied in 8 patients (complete AV block occurred in 1 of 8 patients after 6 months), while ADO II AS was used in 4 patients (no rhythm disturbances in the follow-up). In the latter group, 2 children had pmVSD, one muscular VSD and one residual postsurgical VSD. Three of these patients were closed from the arterial side (similarly as in our material). A residual shunt was observed in one of these 4 patients in the follow-up. In our material, an insignificant residual shunt was seen in 2 out of 6 children. The reason for this phenomenon may be the lack of patches inside the device. In our series as well as in the cited publications, no complications related to the use of ADO II AS were observed, including death, embolization, malposition, haemolysis, thromboembolism, infective endocarditis or vascular problems.
Iatrogenic complete heart block requiring pacemaker implantation remains the most important issue related to transcatheter pmVSD closure and is caused by the proximity of the bundle of His to the border of the defect [4]. After the initial alarming number of cases requiring pacemaker implantation and the subsequent reports of late-onset advanced heart blocks after VSD closure with Amplatzer Perimembranous VSD Occluders (asymmetric) [9], some new occluders were applied as a potential solution. Szkutnik et al. [5] advocated the use of Amplatzer Muscular VSD Occluders in pmVSDs with an aortic rim > 4 mm (length of the skirt of the Muscular VSD Amplatzer Occluder). El Said et al. [10] reported successful application of ADO I in 19 of 21 approached patients with aneurysm-type pmVSD, with no complete heart block; however, in 2 patients the procedure was discontinued due to the occurrence of periprocedural block. ADO I has a mushroom shape, can be implanted only from the venous side and requires a bigger delivery system. It can be particularly useful in patients with pmVSD and a coexisting large aneurysm [11, 12]. Other authors also reported a promising safety profile of Amplatzer Duct Occluder II implantation [13, 14]. Based on a retrospective multicenter registry, Haas et al. presented short and midterm results of perimembranous (81 patients) and muscular (30 patients) VSD closure with Nit-Occlud, proving its high feasibility and low risk of severe adverse effects [15]. Patients with different anatomical variants of pmVSD and psVSD could benefit from variously constructed occluders. A consensus on optimal device selection has not been achieved so far. ADO II AS, as demonstrated by our evidence, could complement the current armamentarium. The disadvantage of this device is that its application is limited to moderate size VSDs (2–4 mm). The retrospective character as well as the small number of patients are the main limitations of our study.

Conclusions

In selected patients with perimembranous and post-surgical VSD the transcatheter application of ADO II AS seems to be safe and effective. However, further data on a larger patient population and long-term follow up are necessary.

Conflict of interest

The authors declare no conflict of interest.

References

1. Scully BB, Morales DL, Zafar F, et al. Current expectations for surgical repair of isolated ventricular septal defects. Ann Thorac Surg 2010; 89: 544-51.
2. Saurav A, Kaushik M, Mahesh Alla V, et al. Comparison of percutaneous device closure versus surgical closure of peri-membranous ventricular septal defects: a systematic review and meta-analysis. Catheter Cardiovasc Interv 2015; 86: 1048-56.
3. Zhou W, Li F, Fu L, et al. Clinical experience of transcatheter closure for residual ventricular septal defect in pediatric patients. Congenit Heart Dis 2016; 11: 323-31.
4. Carminati M, Butera G, Chessa M, et al. Transcatheter closure of congenital ventricular septal defects: results of the European Registry. Eur Heart J 2007; 28: 2361-8.
5. Szkutnik M, Qureshi SA, Kusa J, et al. Use of the Amplatzer muscular ventricular septal defect occluder for closure of perimembranous ventricular septal defects. Heart 2007; 93: 355-8.
6. Champaneri B, Kappanayil M, Kumar RK. Transcatheter closure of membranous septal defect in a 1,8-kg infant using Amplatzer Duct Occluder II additional size device. Cardiol Young 2017; 27: 1437-40.
7. Narin N, Tuncay A, Baykan A, et al. Percutaneous ventricular septal defect closure in patients under 1 year of age. Pediatr Cardiol 2018; 39: 1009-15.
8. Fiszer R, Chojnicki M, Szkutnik M, et al. Are the AMPLATZER duct occluder II additional sizes devices dedicated only for smaller children? EuroIntervention 2017; 12: 2100-3.
9. Walsh MA, Bialkowski J, Szkutnik M, et al. Atrioventricular block after transcatheter closure of perimembranous ventricular septal defects. Heart 2006; 92: 1295-7.
10. El Said, Howaida G, Bratincsak A, et al. Closure of perimembranous ventricular septal defects with aneurysmal tissue using the Amplazter duct occluder I: lessons learned and medium term follow up. Cathet Cardiovasc Interv 2012; 80: 895-903.
11. El-Sisi A, Sobhy R, Jaccoub V, Hamza H. Perimembranous ventricular septal defect device closure: choosing between Amplatzer duct occluder I and II. Pediatr Cardiol 2017; 38: 596-602.
12. Wierzyk A, Szkutnik M, Fiszer R, et al. Transcatheter closure of ventricular septal defects with nitinol wire occluders of type patent ductus arteriosus. Adv Intev Cardiol 2014; 10: 21-5.
13. Polat TB, Turkmen E. Transcatheter closure of ventricular septal defects using the Amplatzer duct occluder II device: a single-center experience. Adv Interv Cardiol 2016; 12: 340-7.
14. Pamukcu O, Narin N, Baykan A, et al. Mid-term results of percutaneous ventricular septal defect closure with Amplatzer duct occluder-II in children. Cardiol Young 2017; 27: 1726-31.
15. Haas NA, Kock L, Bertram H, et al. Interventional VSD-closure with the Nit-Occlud® Lê VSD-Coil in 110 patients: early and midterm results of the EUREVECO-registry. Pediatr Cardiol 2017; 38: 215-27.
Copyright: © 2018 Termedia Sp. z o. o. This is an Open Access article distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 International (CC BY-NC-SA 4.0) License (http://creativecommons.org/licenses/by-nc-sa/4.0/), allowing third parties to copy and redistribute the material in any medium or format and to remix, transform, and build upon the material, provided the original work is properly cited and states its license.
Quick links
© 2024 Termedia Sp. z o.o.
Developed by Bentus.