Original paper

A comparative cost-benefit analysis of electronic brachytherapy vs. high-dose-rate iridium-192 for exclusive vaginal cuff treatment in post-operative endometrial cancer

Sergio Lozares-Cordero
1, 2
,
Víctor González-Pérez
3
,
Marta Sánchez-Casi
1
,
Tomás González-González
1
,
Paula Castillo Peña
1
,
Mireia Pallarés-Ripollés
3
,
Antonio Herreros-Martínez
4
,
Santiago Pellejero-Pellejero
5

  1. Physics and Radiation Protection Department, Miguel Servet University Hospital, Zaragoza, Spain
  2. Department of Microbiology, Pediatrics, Radiology, and Public Health, Faculty of Medicine, University of Zaragoza, Zaragoza, Spain
  3. Medical Physics Department, IVO Foundation, affiliated with the Instituto Valenciano de Oncología, Valencia, Spain
  4. Medical Physics Department, Clinic Hospital, Barcelona, Spain
  5. Medical Physics and Radiation Protection Department, University Complex of Navarra, Pamplona, Spain
J Contemp Brachytherapy 2025; 17, 3: 160–167
Online publish date: 2025/06/30
Article file
Get citation
 
 

Purpose

Endometrial cancer ranks among the most prevalent gynecologic malignancies worldwide, with its incidence rising along with increasing life expectancy, obesity, and metabolic disorders. According to the GLOBOCAN 2018 estimates, it remains a significant cause of cancer-related morbidity and mortality in developed regions [1].

Standard management typically involves total hysterectomy with bilateral salpingo-oophorectomy, occasionally together with lymph node assessment, followed by radiotherapy in selected cases to reduce loco-regional recurrence [2, 3]. For patients with low- to intermediate-risk disease, post-operative vaginal brachytherapy demonstrates excellent local control while maintaining favorable toxicity profiles [4, 5].

High-dose-rate (HDR) brachytherapy using an iridium-192 (192Ir) source has traditionally been the modality of choice, supported by robust clinical data, well-established protocols, and widespread availability [6]. However, technical advances have led to the development of electronic brachytherapy (eBT) that produces low-energy X-rays in situ. This technology avoids the need for transporting, storing, and disposing of radioactive isotopes, potentially offering advantages in terms of radiation safety and infrastructure requirements [7, 8]. While eBT has gained traction in early-stage breast cancer and dermatologic indications, its application in gynecologic malignancies has been limited [9]. Early evidence, however, suggests comparable dosimetric and clinical outcomes between eBT and 192Ir HDR for endometrial cancer, prompting interest in its broader adoption [10, 11]. From an economic perspective, each approach incurs distinct costs and benefits. Conventional HDR brachytherapy with 192Ir may require dedicated bunkers, specialized source handling, and periodic source replacements, thus incurring ongoing expenses [12]. In contrast, eBT involves a higher upfront equipment cost but eliminates radioactive source replenishment, and may allow treatments in rooms with less extensive shielding [7, 9]. Where clinical effectiveness appears similar, it becomes critical to assess cost differentials, factoring in setup expenses, maintenance contracts, consumable usage, and potential logistical efficiency [13]. Moreover, as healthcare systems increasingly focus on value-based care, delivering high-quality treatments at sustainable costs remains a top priority [14].

In this study, we presented a cost-benefit analysis comparing exclusive 192Ir brachytherapy and eBT in patients with endometrial cancer. Our primary goal was to quantify the financial implications of each modality, encompassing acquisition, maintenance, and procedural expenses under the assumption of equivalent oncologic outcomes. By highlighting the key drivers of the costs and possible areas for optimization, we aimed to provide clinicians, hospital administrators, and policy-makers with evidence-based insights to guide technology selection and resource allocation. Ultimately, we hoped this analysis would inform strategies to maintain or enhance clinical efficacy while minimizing economic burdens for a growing population of patients requiring radiotherapeutic management of endometrial cancer [11].

In addition to the cost-benefit evaluation, this study provided a detailed dosimetric comparison between HDR 192Ir and electronic brachytherapy modalities. Although previous investigations have indicated comparable dosimetric profiles [11], such an analysis would strengthen the clinical relevance of the findings, demonstrating more holistic approach to technology selection in the management of post-operative endometrial cancer.

Material and methods

This retrospective observational study analyzed the cost-benefit of exclusive brachytherapy in endometrial cancer patients treated in two Spanish centers between 2019 and 2023. All patients included in this analysis had completed treatment before the evaluation period. The primary objective was to compare the economic and dosimetric outcomes of eBT and HDR 192Ir brachytherapy, specifically focusing on patients treated exclusively with brachytherapy and without supplementary external beam radiotherapy (EBRT), to maintain consistency in the treatment modality. One center is equipped with HDR 192Ir unit, while the other uses eBT technology.

A total of 70 patients were treated with 192Ir HDR brachytherapy, and 115 patients received eBT. Inclusion criteria comprised type I FIGO stage IA endometrial cancer with risk factors, type I FIGO stage IB, and, especially, type II FIGO stage II in patients whose comorbidities precluded EBRT (Table 1). Baseline clinical and tumor characteristics (i.e., FIGO stage, histologic grade, and lymphovascular invasion) were compared between eBT and 192Ir groups.

Table 1

Tumor characteristics in patients treated with electronic brachytherapy (eBT) vs. high-doserate (HDR) iridium-192 (192Ir)

Characteristic192Ir (n = 70)eBT (n = 115)
Endometrioid histology (type I)88%89%
Lymphovascular invasion19%15%
Moderately differentiated (grade 2)60%65%
Myometrial invasion > 50%27%25%
FIGO stage IA78%69%
FIGO stage IB21%29%
FIGO stage II1%2%

Regarding brachytherapy planning and delivery for vaginal cuff irradiation, each patient received 3 fractions of 7 Gy, yielding a total dose of 21 Gy prescribed at a 5 mm depth from the applicator surface, in accordance with established clinical guidelines for adjuvant treatment in endometrial cancer [4]. Two treatment modalities were employed. In the 192Ir HDR approach, treatments were planned using Oncentra Brachy system (version 4.6.3, Elekta, Stockholm, Sweden), and a radioactive 192Ir source with a half-life of approximately 74 days was utilized, undergoing quarterly replacements. The center, which uses HDR 192Ir equipment, is treating an average of over 220 patients annually, performing more than 740 sessions across 17 different locations. Alternatively, eBT was delivered using BrachyVision system (Varian Medical Systems, Palo Alto, CA), coupled with a 50 kVp X-ray source (Xoft Axxent). The eBT unit produces low-energy X-rays, enabling treatments to be delivered in minimally shielded procedure rooms, thereby eliminating the need for specialized radiation shielding.

Staff resources required for both treatment modalities, HDR 192Ir and eBT, are quite similar. Both modalities require trained professionals for patient setup, treatment delivery, and monitoring during the procedure. Tasks performed by medical physicists, radiation oncologists, and other clinical staff are comparable in both approaches, as both involve precision placement and radiation dose delivery, although differing in technology.

Regarding commissioning and QA time, the procedures are also similar. Both modalities require initial commissioning and regular QA to ensure optimal treatment delivery. Processes involved in both types of brachytherapy are analogous, including source calibration, verification of treatment plans, and system maintenance.

The eBT sources require replacement every 750 minutes of clinical use, corresponding to a cost per patient based on an average treatment duration of approximately 5 minutes per session. Dose optimization was achieved by adjusting dwell times along the vaginal cylinder to ensure homogeneous coverage. The cost of each source replacement was included in the cost per minute (the source has an autonomy of 750 minutes, therefore its cost is divided by the total number of minutes and multiplied by the minutes used for the endometrial cases analyzed) of treatment in the case of eBT [4]. Prior to treatment, patients underwent computed tomography (CT) simulation in a supine position with a slice thickness of 2-3 mm. CT images facilitated precise delineation of clinical target volume (CTV) defined as the upper 3-5 cm of the vaginal cuff and of organs at risk (OARs), such as the bladder, rectum, and sigmoid colon (Table 2) [15]. Dose distributions were optimized to deliver 7 Gy per fraction to 5 mm from the CTV while minimizing exposure to OARs, with key dosimetric parameters, including D2cc, D1cc, and D0.1cc for the bladder, rectum, and sigmoid systematically recorded (Table 2). Each treatment plan was reviewed by a radiation oncologist and a medical physicist to ensure adherence to dose constraints [16, 17]. Moreover, comprehensive quality assurance procedures were implemented. For HDR 192Ir brachytherapy, source calibration, dwell times, dwell position tests, and periodic activity checks following each source replacement were verified, whereas for eBT, X-ray source output was confirmed using calibrated detectors prior to each treatment fraction [18].

Table 2

Planning parameters for vaginal cuff brachytherapy

ParameterCenter A (192Ir)Center B (eBT)
CT slice thickness (mm)2-32-3
Applicator diameters (cm)2.5/3/3.52.5/3/3.5
Prescription depth from applicator (mm)55
Number of fractions33
Dose per fraction (Gy)77
Total prescribed dose (Gy)2121
Planning softwareOncentra BrachyBrachyVision

An activity-based cost analysis was conducted from the hospital’s perspective to determine the total cost of each brachytherapy modality by incorporating all relevant expense components across the centers. Capital equipment expenditures, including acquisition of brachytherapy units (HDR afterloader and eBT generator) were annualized over their respective lifespans using a straight-line depreciation model, with one approach assuming an eight-year lifespan and another allocating depreciation over a standard equipment lifespan of five to ten years in accordance with center-specific financial policies. Source replacement and maintenance costs were subsequently evaluated, with HDR 192Ir necessitating quarterly radioactive source replacement, encompassing procurement, shipping, and disposal fees, while eBT requiring annual service contracts for the electronic source, source replacement every 750 minutes of clinical use, and routine maintenance. Overhead and infrastructure costs, such as differential shielding requirements between HDR 192Ir and eBT procedures, were not included in the analysis due to variability in local construction costs and broader clinical use of 192Ir bunkers (as discussed later in the manuscript), whereas eBT procedures can be performed in standard rooms with minimal additional shielding [12]. Consumable costs per session, with various items, such as sterile applicator covers and lubricants, were estimated, with ancillary operational costs (e.g., electricity consumption and minor disposables) assumed equivalent between both techniques, unless center-specific data indicated otherwise. Personnel costs calculated based on standardized salary data from the Spanish National Health System were considered identical for both modalities, given the consistent treatment schedule (3 fractions of 7 Gy) and utilization of equivalent staff resources, thereby negating additional labor cost differentials. Global cost figures for each center were derived by summing annualized equipment depreciation, source replacement and maintenance, infrastructure, and consumable costs, and then normalizing these totals to the overall number of brachytherapy sessions performed annually across departments in all brachytherapy procedures. The resulting per-session cost was multiplied by 3 to yield the cost per patient. Both modalities were implemented using established systems. The economic evaluation of this study focused exclusively on cost differences associated with different equipment used in each modality. Specifically, both techniques involved the same number of medical consultations, equivalent time allocation for medical physics assessments and quality control, and identical physician supervision, ensuring uniformity in all procedural aspects.

Additionally, vaginal cylinders of various diameters were utilized to accommodate patient anatomy, with applicators sterilized and prepared per each institution’s standard operating procedures. This comprehensive cost evaluation integrated capital investment, operational expenditures, and personnel costs, to provide a detailed economic assessment of HDR 192Ir vs. eBT, thereby demonstrating resource allocation and decision-making processes in clinical practice [19, 20].

Results

The costs were analyzed across several categories, including staff, equipment, and operating expenses, with values provided in both EUR and USD. The key cost components, such as acquisition and depreciation, source replacement and maintenance as well as infrastructure and quality assurance, are summarized in Table 3.

Table 3

Cost comparison between high-dose-rate (HDR) iridium-192 (192Ir) and electronic brachytherapy (eBT) for exclusive endometrial brachytherapy (applicable to Spain under current economic conditions and assumptions of this work, with exchange rate as of March 2025: 1 EUR = 1.1160 USD)

Cost item/parameter192Ir (EUR)192Ir (USD)eBT (EUR)eBT (USD)Difference (eBT vs. 192Ir) (EUR)Difference (eBT vs. 192Ir) (USD)
I. Staff costs (per patient)1330.0368.3330.0368.30.00.0
II. Equipment costs
A) Acquisition and depreciation (per patient)2344.5384.5299.5334.2–45–50.3
B) Source replacement and maintenance (per patient)565.2631.4336.3375.6–228.9–255.8
C) Infrastructure, TPS, and QA (per patient)100.0111.6100.0111.60.00.0
III. Operating costs
Consumables total (3 sessions) (per patient)3150.0167.4150.0167.40.00.0
IV. Estimated total cost under assumptions of this work (per patient)41,489.71,662.51,215.81,356.8–273.9 (–18.4%)–305.7 (–18.4%)

1 Identical staff (medical consultations, physics support, and physician supervision) for both modalities; 2 Acquisition and depreciation (per patient) refers to equipment acquisition and depreciation costs calculated based on clinical usage per patient. This category includes the initial investment made when purchasing the equipment (either HDR 192Ir or eBT), amortized over its expected lifespan and clinical utilization frequency, resulting in an estimated per-patient cost reflecting equipment depreciation; 3 Covers all clinical consumables, except for the radiation source (applicators, covers, lubricants, etc.), assumed identical for both HDR 192Ir and eBT; 4 Sum of staff costs (I) + equipment costs (II) + operating costs (III); *Note – personnel costs are based on average salary data from the Spanish National Health System [21]. Costs are applicable to Spain under current economic conditions

The primary economic differentiators between the two techniques were the costs associated with equipment usage, including source replacement and maintenance expenses. The distribution of total sessions and exclusive endometrial brachytherapy sessions per year for both modalities is presented in Table 4, with the cost per session derived from source replacement and maintenance expenses.

Table 4

Annual distribution of brachytherapy sessions for iridium-192 (192Ir) and electronic brachytherapy (eBT)

192Ir20192020202120222023
Total patient sessions646741767706848
Exclusive endometrial patients1216111516
Exclusive endometrial sessions3648334548
eBT20192020202120222023
Total patient sessions361375377354388
Exclusive endometrial patients2025241927
Exclusive endometrial sessions6075725781

For the purpose of this cost-benefit evaluation, we assumed there would be no significant differences in dosimetric outcomes between eBT and 192Ir HDR brachytherapy utilization in patients with endometrial cancer [11]. This assumption was supported by studies indicating that eBT provides comparable efficacy and safety profiles to HDR 192Ir brachytherapy for vaginal cuff irradiation, with similar rates of local recurrence and treatment-related toxicities [7, 9]. As such, any cost differentials observed in this analysis primarily reflected differences in equipment, source replacement, and maintenance, rather than variations in clinical efficacy or complication rates.

The dosimetric comparison between eBT and 192Ir HDR brachytherapy is summarized in Table 5. All values are expressed as a percentage of the prescribed dose (7 Gy per fraction), while mean values, standard deviations (SD), and ranges are presented for each organ at risk (OAR) parameter, reflecting the distribution of doses across all patients in each group, with independent samples t-test employed.

Table 5

Dosimetric comparison between electronic brachytherapy (eBT) and HDR 192Ir (as % of 7 Gy prescribed dose)

Parameter192Ir, mean ±SD (range)eBT, mean ±SD (range)p-value
CTV D90%101.2 ±2.3 (98-105)100.7 ±2.5 (97-104)0.487
CTV V150%31.4 ±3.9 (24-38)36.1 ±4.3 (29-44)0.063
CTV V200%10.8 ±2.5 (7-15)14.2 ±3.0 (9-19)0.071
Bladder D2cc69.8 ±4.7 (61-78)65.5 ±4.2 (58-73)0.112
Bladder D1cc72.0 ±3.5 (64-80)71.0 ±3.8 (63-79)0.675
Bladder D0.1cc74.5 ±3.9 (67-82)78.2 ±4.5 (70-87)0.205
Rectum D2cc67.9 ±4.3 (60-76)64.8 ±4.1 (57-72)0.245
Rectum D1cc70.1 ±3.8 (63-77)70.5 ±4.0 (63-78)0.782
Rectum D0.1cc72.8 ±4.0 (65-81)74.7 ±4.2 (67-83)0.301
Sigmoid D2cc62.9 ±4.2 (55-70)59.8 ±4.1 (52-67)0.158
Sigmoid D1cc68.2 ±3.9 (60-75)68.4 ±4.0 (60-76)0.894
Sigmoid D0.1cc71.6 ±4.1 (64-79)72.4 ±4.0 (65-80)0.725

Overall, the target coverage was comparable between the two modalities (data not shown), and OARs doses demonstrated slight variations favoring one technique or the other, depending on the specific parameter. For instance, eBT yielded lower mean bladder D2cc, but slightly higher bladder D0.1cc, whereas rectal and sigmoid parameters exhibited similar trends between eBT and 192Ir HDR, but not statistically significant.

These dosimetric findings align with previously published reports [11], suggesting that eBT can achieve dose distributions comparable to 192Ir HDR brachytherapy, while maintaining acceptable OARs constraints for post-operative endometrial cancer treatments.

Discussion

The results of this study confirm that eBT offers a viable alternative for treating post-operative endometrial cancer, particularly in centers without access to 192Ir HDR brachytherapy. While 192Ir remains the gold standard due to its flexibility in treating multiple anatomical sites, eBT presents unique advantages, including reduced shielding requirements, mobility, and lower maintenance costs associated with source handling. These attributes align with findings from previous studies, which highlight the practicality of eBT in centers with limited infrastructure or in low-resource settings [12, 20]. Additionally, eBT can represent a valuable complementary option for centers performing brachytherapy treatments with high volumes. However, it should be noted that eBT cannot be utilized for interstitial brachytherapy procedures due to larger diameter of its source.

Importantly, this study does not suggest replacing 192Ir with eBT, but rather identifies its complementary role. By selectively allocating eBT for endometrial cancer as well as other indications, including non-melanoma skin cancer and intraoperative radiotherapy for breast cancer, 192Ir can be reserved for more complex treatments, such as cervical cancer, requiring interstitial components [22, 23]. This approach optimizes resource allocation and ensures that the broad spectrum of 192Ir applications is maintained while expanding access to brachytherapy for less technically demanding cases.

We acknowledge that the construction cost of a dedicated brachytherapy bunker for 192Ir treatments could be allocated on a per-patient basis. For instance, assuming a construction cost ranging between €500,000 and €1,000,000 and amortizing this investment over 30 years with an annual patient volume of 200, the resulting per-patient cost would be estimated between approximately €83 and €166, respectively. However, this cost was not directly incorporated into our economic analysis for several reasons. First, there is substantial variability in bunker construction costs due to differences in local market conditions and contractor pricing, which makes a universal estimate challenging. Second, while eBT does not include bunker-related expenses, 192Ir brachytherapy is often employed across a broader spectrum of clinical applications beyond endometrial cancer. This wider utilization allows facilities to distribute the bunker cost over a larger patient population, further diluting the per-patient expense. Given that our study specifically focused on endometrial treatments, we opted not to directly add this variable cost to each patient’s result, thereby ensuring that our economic evaluation remains both focused and reflective of real-world clinical practice.

Considering the number of sessions performed annually (740 on average) at the analyzed 192Ir center, each additional endometrial patient treated would reduce the cost per-patient by approximately 0.5% of the equipment cost (Table 3), provided there is sufficient time to accommodate them, and considering other patient locations. Conversely, with eBT, each additional endometrial treatment would increase the cost by approximately 0.08%, which could lead to a convergence of both modalities in terms of cost per patient when treating a sufficiently high number of cases. However, prioritizing endometrial treatments with 192Ir would influence the availability of the system for other localizations, typically more demanding in terms of professional time, thus reducing the overall treatment capacity and flexibility of the HDR unit. Additionally, at the 192Ir center, the maximum number of treatments is already being performed with the available staff, making it unfeasible to increase the number of patients. Therefore, in this scenario, extrapolating costs as a function of the number of patients would be difficult. Although, the relative cost per patient decreases with more patients in the 192Ir scenario, but increases in the eBT scenario, due to the different allocation of source-related costs per irradiation minute. This underscores the strategic advantage of complementing the 192Ir system with eBT in such situations, preserving versatility while optimizing resource allocation.

Dosimetric outcomes in this study align with previous research indicating that eBT achieves comparable CTV coverage to 192Ir while reducing doses to OARs, such as the bladder and rectum [8, 11, 23]. This is particularly relevant in demonstrating its safety profile, as observed in studies where acute and long-term toxicities were minimal, even in higher dose regions, such as the vaginal mucosa [23]. Clinicians must weigh these factors when selecting the appropriate modality to optimize patient outcomes.

Economic analyses also support the feasibility of eBT, particularly when considering savings associated with eliminating radioactive source handling (eBT source replacements are less costly and logistically simpler compared with radioactive source replacements) and minimizing logistical burdens. Previous studies demonstrated how reducing fractions in vaginal cuff brachytherapy significantly decreased costs without compromising clinical outcomes [8]. Similar savings are evident when eBT is incorporated for exclusive endometrial treatments, leveraging its mobility and reduced infrastructure demands [24].

Conversely, HDR 192Ir brachytherapy systems have lower upfront costs, but incur ongoing expenses related to source replacement and rigorous safety protocols. A longer half-life of alternative isotopes, such as cobalt-60 (60Co) was proposed as a cost-effective option, reducing the frequency of source replacements. However, 60Co presents its own set of dosimetric [25] and logistical considerations [26]. Dosimetrically, 60Co and 192Ir sources are comparable in clinical applications, with studies indicating negligible differences in treatment outcomes [27]. However, 60Co’s higher energy emissions necessitate additional shielding considerations in facility design and may require specific regulatory approvals, potentially impacting initial setup costs and operational logistics. In terms of costs, while 60Co sources may have a higher upfront acquisition costs, the extended replacement interval can lead to lower cumulative expenses over time. This cost dynamic is particularly beneficial for centers with limited resources, or those aiming to minimize frequent source procurement and associated logistical challenges.

This study offers a comprehensive evaluation of both the dosimetric and economic aspects of eBT and HDR 192Ir brachytherapy, making it the only investigation to directly compare economic aspects of these modalities in post-operative treatment of endometrial cancer. The inclusion of 185 patients from two centers ensures a robust dataset, improving the reliability of the findings. Furthermore, the detailed dosimetric data, with parameters, such as D2cc, D1cc, and D0.1cc for OARs, offer valuable insights into the clinical implications of each modality. The economic analysis, which calculates costs per session while considering both source replacement and maintenance expenses, provides an additional layer of practicality to the study.

Despite its strengths, the current research has several limitations, which should be acknowledged. First, the analysis excludes certain indirect costs, such as those related to personnel training, infrastructure upgrades, and long-term maintenance beyond the evaluation period. Second, the study does not include clinical follow-up data on toxicities or oncologic outcomes, which limits the ability to correlate dosimetric and economic findings with patient quality of life or disease control. Finally, the generalizability of these results may be constrained by the specific settings of the two centers included, as institutional practices and cost structures can vary significantly. Additionally, the findings may not be directly applicable to other countries or centers with different workloads or operational conditions, where the volume of treatments and available resources could influence the cost-efficiency observed in this study.

These limitations highlight the need for caution when applying the findings to other clinical settings, emphasizing the importance of individualized decision-making based on local resources and patient populations.

It should be noted that HDR brachytherapy sources, such as 192Ir, deplete continuously due to radioactive decay regardless of patient throughput. Consequently, healthcare centers already having HDR brachytherapy afterloaders can potentially improve resource efficiency and economic sustainability by increasing the utilization of these units for additional clinical indications, such as interstitial prostate brachytherapy, rather than automatically defaulting EBRT. This strategic approach to brachytherapy resource management could lead to better optimization of source usage, thus enhancing cost-efficiency and potentially offsetting the apparent economic advantages identified in eBT compared with HDR brachytherapy.

This consideration might significantly influence the cost-benefit conclusions derived from the present analysis. Specifically, if the HDR afterloader equipment is utilized for multiple anatomical sites or diverse clinical indications, the cost per patient attributed to source replacement and maintenance decreases substantially due to broader resource allocation. Therefore, the calculated economic differences favoring eBT in scenarios strictly limited to endometrial cancer treatments could be narrowed or even reversed, if broader clinical applications of HDR brachytherapy are considered.

Additionally, the relatively low cost of treating endometrial cancer patients could support the complementary usage of eBT in centers equipped with both technologies. This would allow healthcare providers to efficiently allocate resources, utilizing HDR for more complex or resource-intensive procedures while reserving eBT for less technically demanding treatments, thereby optimizing overall cost-effectiveness and operational flexibility.

Future research should focus on integrating clinical outcomes with dosimetric and economic analyses to provide a more comprehensive understanding of long-term implications of eBT and 192Ir HDR brachytherapy. Studies exploring the impact of eBT on acute and late toxicities, particularly in high-dose regions, such as the vaginal mucosa, are warranted to further validate its safety profile. Moreover, comparative studies involving newer technologies and isotopes, such as 60Co or advanced eBT systems, could shed light on potential improvements in cost-effectiveness and clinical performance.

Technological advancements may also influence the cost structure and clinical adoption of these modalities. Innovations in source design, applicator systems, and treatment planning algorithms can enhance the efficiency and precision of eBT and 192Ir HDR, potentially narrowing the economic gap between the two techniques. Moreover, expanding the indications for eBT beyond endometrial cancer to include other gynecologic or non-gynecologic malignancies might improve its cost-effectiveness through higher utilization rates.

Conclusions

This study demonstrates that both eBT and HDR 192Ir brachytherapy achieve comparable dosimetric outcomes in the exclusive treatment of the vaginal cuff following surgery for endometrial cancer. While the allocation of costs to endometrial cylinder treatments remains complex due to the shared use of HDR equipment for a variety of indications, eBT presents distinct advantages by reducing the relative recurrent costs (based on Spanish pricing) associated with radioactive source management and by lessening infrastructure demands.

With combining economic and dosimetric evaluations, this work offers a comprehensive framework of technology selection, advocating for a complementary model, in which eBT is prioritized for less complex cases and HDR 192Ir is reserved for broader or more demanding clinical applications, ultimately optimizing resource use and patient outcomes.

Funding

This research received no external funding.

Disclosures

Approval of the Bioethics Committee was not required.

Notes

[2] Conflicts of interest The authors declare no conflict of interest.

References

1 

Bray F, Ferlay J, Soerjomataram I et al. Global cancer statistics 2018: GLOBOCAN estimates of incidence and mortality worldwide for 36 cancers in 185 countries. CA Cancer J Clin 2018; 68: 394-424.

2 

Morice P, Leary A, Creutzberg C et al. Endometrial cancer. Lancet (London, England) 2016; 387: 1094-1108.

3 

Creutzberg CL, Nout RA, Lybeert MLM et al. Fifteen-year radiotherapy outcomes of the randomized PORTEC-1 trial for endometrial carcinoma. Int J Radiat Oncol Biol Phys 2011; 81: e631-638.

4 

Nout RA, Smit VTHBM, Putter H et al. Vaginal brachytherapy versus pelvic external beam radiotherapy for patients with endometrial cancer of high-intermediate risk (PORTEC-2): an open-label, non-inferiority, randomised trial. Lancet (London, England) 2010; 375: 816-823.

5 

Sorbe B, Horvath G, Andersson H et al. External pelvic and vaginal irradiation versus vaginal irradiation alone as postoperative therapy in medium-risk endometrial carcinoma–a prospective randomized study. Int J Radiat Oncol Biol Phys 2012; 82: 1249-1255.

6 

Harkenrider MM, Block AM, Alektiar KM et al. American Brachytherapy Task Group Report: Adjuvant vaginal brachytherapy for early-stage endometrial cancer: A comprehensive review. Brachytherapy 2017; 16: 95-108.

7 

Dickler A, Puthawala MY, Thropay JP et al. Prospective multi-center trial utilizing electronic brachytherapy for the treatment of endometrial cancer. Radiat Oncol 2010; 5: 67.

8 

Rivard MJ, Davis SD, DeWerd LA et al. Calculated and measured brachytherapy dosimetry parameters in water for the Xoft Axxent X-Ray Source: an electronic brachytherapy source. Med Phys 2006; 33: 4020-4032.

9 

Kamrava M, Chung MP, Demarco J et al. Electronic brachytherapy for postsurgical adjuvant vaginal cuff irradiation therapy in endometrial and cervical cancer: a retrospective study. Brachytherapy 2013; 12: 141-147.

10 

Dickler A, Kirk MC, Coon A et al. A dosimetric comparison of Xoft Axxent Electronic Brachytherapy and iridium-192 high-dose-rate brachytherapy in the treatment of endometrial cancer. Brachytherapy 2008; 7: 351-354.

11 

Lozares-Cordero S, Font-Gómez JA, Gandía-Martínez A et al. Postoperative endometrial cancer treatments with electronic brachytherapy source. J Radiother Pract 2019; 18.

12 

Mobit PN, Nguyen A, Packianathan S et al. Dosimetric comparison of brachytherapy sources for high-dose-rate treatment of endometrial cancer: 192Ir, 60Co and an electronic brachytherapy source. Br J Radiol 2016; 89: 20150449.

13 

Ahmad S, Johnson D, Hiatt JR et al. Comparison of tumor and normal tissue dose for accelerated partial breast irradiation using an electronic brachytherapy eBx source and an Iridium-192 source. J Appl Clin Med Phys 2010; 11: 155-161.

14 

Beaulieu L, Carlsson Tedgren Å, Carrier JF et al. Report of the Task Group 186 on model-based dose calculation methods in brachytherapy beyond the TG-43 formalism: Current status and recommendations for clinical implementation. Med Phys 2012; 39: 6208-6236.

15 

Haie-Meder C, Pötter R, Van Limbergen E et al. Recommendations from Gynaecological (GYN) GEC-ESTRO Working Group (I): concepts and terms in 3D image based 3D treatment planning in cervix cancer brachytherapy with emphasis on MRI assessment of GTV and CTV. Radiother Oncol J Eur Soc Ther Radiol Oncol 2005; 74: 235-245.

16 

Kamrava M, Leung E, Bachand F et al. GEC-ESTRO (ACROP)-ABS-CBG Consensus brachytherapy target definition guidelines for recurrent endometrial and cervical tumors in the vagina. Int J Radiat Oncol Biol Phys 2023; 115: 654-663.

17 

Wortman BG, Astreinidou E, Laman MS et al. Brachytherapy quality assurance in the PORTEC-4a trial for molecular-integrated risk profile guided adjuvant treatment of endometrial cancer. Radiother Oncol J Eur Soc Ther Radiol Oncol 2021; 155: 160-166.

18 

Soror T, Siebert FA, Lancellotta V et al. Quality assurance in modern gynecological HDR-brachytherapy (interventional radiotherapy): Clinical considerations and comments. Cancers (Basel) 2021; 13: 912.

19 

Vu CC, Blas KG, Lanni TB et al. Cost-effectiveness of prostate boost with high-dose-rate brachytherapy versus intensity-modulated radiation therapy in the treatment of intermediate-high risk prostate cancer. Brachytherapy 2018; 17: 852-857.

20 

Zhang Y, Rovirosa A, Ascaso C et al. Economic impact of decreasing the fraction number in vaginal cuff brachytherapy: A direct cost analysis. Brachytherapy 2020; 19: 60-65.

21 

Ministerio de Sanidad, Consumo y Bienestar Social. Informe sobre costes y financiación del Sistema Nacional de Salud. Madrid, Spain; 2022 (accessed February 2025). n.d. https://www.sanidad.gob.es/estadEstudios/estadisticas/sisInfSanSNS/tablasEstadisticas/InfAnSNS.htm.

22 

Lozares-Cordero S, Font-Gómez JA, Gandía-Martínez A et al. Treatment of cervical cancer with electronic brachytherapy. J Appl Clin Med Phys 2019; 20: 78-86.

23 

Hitova-Topkarova D, Payakova V, Kostova-Lefterova D et al. Electronic brachytherapy for gynecological cancers–a systematic review. Reports Pract Oncol Radiother J Gt Cancer Cent Pozn Polish Soc Radiat Oncol 2023; 28: 79-87.

24 

Lozares-Cordero S, Gonzalez-Perez V, Pellejero-Pellejero S et al. Feasibility of electronic brachytherapy in cervix cancer–a dosimetric comparison of different brachytherapy techniques. Brachytherapy 2022; 21: 389-396.

25 

Palmer A, Hayman O, Muscat S. Treatment planning study of the 3D dosimetric differences between Co-60 and Ir-192 sources in high dose rate (HDR) brachytherapy for cervix cancer. J Contemp Brachytherapy 2012; 4: 52-59.

26 

Vega RM, Duckworth T, DeWyngaert JK et al. Cost-benefit analysis of Co-60 HDR afterloaders in management of gynecological malignancies: What constitutes an acceptable shielding cost? Brachytherapy 2015; 14: S37-S38.

27 

Granero D, Pérez-Calatayud J, Ballester F. Technical note: Dosimetric study of a new Co-60 source used in brachytherapy. Med Phys 2007; 34: 3485-3488.

Copyright: © 2025 Termedia Sp. z o. o. This is an Open Access article distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 International (CC BY-NC-SA 4.0) License (http://creativecommons.org/licenses/by-nc-sa/4.0/), allowing third parties to copy and redistribute the material in any medium or format and to remix, transform, and build upon the material, provided the original work is properly cited and states its license.
 