Purpose

Locally advanced or recurrent pelvic soft tissue or gynecological tumors can be challenging for radical surgery, especially when abutting the pelvic sidewall. For patients with no previous radiotherapy, pre- or post-operative external beam radiation therapy (EBRT) is usually applied to reduce the risk of local recurrence, but the risk of microscopic involved margins after resection remains high. For patients who recur after previous radiotherapy, the dose is often limited by the risk of exceeding the normal tissue tolerance. Intra-operative external beam radiotherapy (IORT) or brachytherapy (BT) can allow dose escalation that can be used in such situations, delivering a perfectly targeted and localized boost dose in the perioperative period, where microscopic residues are expected, and potentially improving outcome [1,2,3,4]. Permanent low-dose-rate implants with iodine 125 (I-125) seeds, which are mostly used for prostate interstitial BT [5], could have numerous advantages compared to other techniques of perioperative dose delivery in the treatment of deep-seated pelvic tumors [6,7].
We conducted an experimental animal study to test the technical feasibility of this perioperative I-125 brachytherapy mesh implantation in a pelvic location, before applying it clinically.

Material and methods

Animal selection

Animal selection was based on anatomical constraints similar to humans and species with low weight gain. Therefore, we used the following criteria: large adult animals, mobile and constant weight to allow unbiased evaluation during 6-month study. Thus, Romanov ewes supplied by INRA, the French national agricultural research body, aged 3 to 6 years and weighing between 50 and 60 kg were chosen for the study.

Study design and experimental workflow

This experimental study was approved by the ethical committee for animal experimentation in the Midi-Pyrenees region, France. All the partners agreed not to euthanize the animals at the end of the study. They returned to the INRA farm to lead a normal life, in particular for reproductive purposes.
We first implanted non-radioactive dummy seeds similar to the radioactive iodine 125 seeds used in humans, with a bio-compatible flexible sterile absorbable mesh. No ewe became pregnant during the study in order to avoid the displacement of small loops by a pregnant uterus. For the same reason, all meshes were implanted on the right side of the animal, so that the rumen, which may sometimes occupy the entire left abdomen, did not vary in volume. Meshes were therefore positioned by laparotomy in the pelvic area under general left lateral decubitus anesthesia and with minimal manipulation to avoid seed movement (Figure 1). Anesthesia was performed by a veterinary surgeon and surgery by an oncologist surgeon with an animal experimentation license.
The surgical approach was by a right transversal incision and implantation was performed after dissecting the right pelvic peritoneum. The meshes were attached to the underlying muscle wall by atraumatic clips. The intervention time ranged from 40 to 60 minutes per animal, including anesthesia, mowing, shaving, skin preparation, mesh implantation, and abdominal closure in three planes. For this study, 45 dummy seeds (BARD Medical, Covington, USA) were placed evenly on a 10 cm2 polyglactin (Vicryl) mesh (Figure 2). The seeds were positioned so as to obtain a dose of 160 Gy at 5 mm from the center of each seed. Therefore, the seeds were shifted, and the center-to-center distance of the seeds was 10 mm for each line; the diagonal distance between each seed was 1.4 cm (= √2). The seeds were glued on the mesh with a sterile durable cyano-acrylate adhesive solution. This polyglactin mesh can be easily cut to match irregular anatomical areas.
The dummy seeds used in this study were the same as those used in human surgery. They were hermetically sealed in a titanium capsule, medical grade ASTM F67, and the ends were laser-welded. The thickness of the titanium capsule was 0.08 mm and that of the welds was 0.13 mm. Titanium is the only material that encounters the tissues and is perfectly biocompatible. Normally, the radioactive seeds contain iodine 125, which is adsorbed on an aluminum wire with a nickel copper-plated gold core. The gold contained in the wire serves as a radio-opaque X-ray marker, thereby allowing visualization after implantation. The copper layer serves as a physical substrate for the adsorption of iodine 125. The device had the following dimensions: D 0.8 mm × L 4.5 mm (Figure 3).
Finally, after the recovery period, the animals were transferred to the INRA farm and then transported to undergo dosimetric CT scans. Three CT scans were performed on a dedicated veterinary CT scanner at day 15, day 70, and day 180 after implantation. These three CT scans are called CT1, CT2, and CT3, respectively. CT scans were completed under general anesthesia (identical protocol) to limit their stress and avoid movement.

Dosimetric analysis

Seed position and geometric verification during surgery and follow-up in living animals were major issues in this study. For example, if the seeds are spaced irregularly, this may cause significant dose modifications that, in turn, may lead to under-detection or over-dosing, with the foreseeable consequences of inefficient treatment and the risks of complications. A preliminary analysis with an Advantage Workstation (GE, version 4.4, USA) was carried out in order to check the positioning of the plates and any seed migration. Then, the mesh coordinates were determined between the centroid of the plate and the bone anatomical landmarks to have the same mark on the two CT scans at D+15 (CT1) and D+180 (CT3) (Figure 4). The tolerance of a shift was set at 10 mm in all directions (X, Y, Z).
Finally, a dosimetric analysis was performed on VariSeedTM (Treatment Planning System, v8.0.2, Varian Medical Systems, Inc., Palo Alto, CA, USA) between CT1, CT2, and CT3. Seeds were detected automatically on each CT scan. Seed activity was fixed arbitrarily at 1 U (= 1 uGy.m2/h) or 0.787 mCi, to have a 160 Gy dose line around the mesh volume and 5 mm from the center of each seed (Figure 5).

Statistical analysis

The primary endpoint was the feasibility rate. Feasibility of success was defined as a difference in the 160 Gy isodose line volume lower than ±10%. Using a Simon minimax design with α = 10, 1-β = 90, p0 = 0.70 (maximum percentage of non-feasibility) and p1 = 0.95 (minimum percentage of feasibility), 17 ewes were included in the experimental arm. An interim analysis was performed after the inclusion of 10 ewes. If the procedure had a feasibility of success in 7 (or fewer) out of 10 animals, the procedure could be considered insufficiently feasible.
At the end of the second stage, if at least 13 ewes showed a feasibility of success, it could be considered sufficiently feasible for further investigation. Continuous data were presented using median and range. Comparisons between continuous paired data were performed using the Wilcoxon paired test.

Results

In the first phase, one out of 10 animals did not undergo the third scan and was excluded from the study. In total, 16 ewes were analyzed, 9 in the first phase and 7 in the second.

Number of seeds and dose distribution

No seed migration was observed between the three scans; the 45 seeds remaining detectable in the 16 ewes. For 8 of 9 animals, the 160 Gy isodose line volume was within 10%, thereby testifying the feasibility of success and allowing 7 more ewes to be added to the study. The difference in volume of the 160 Gy isodose line between the three scans and for each ewe is shown in Table 1 and Figure 6. All 16 ewes were exploitable, and 13 had a volume difference ≤ 10%. The absolute median volume difference values were 5.64% (range: 0.24-21.03) between CT1 and CT3, 3.07% (range: 0.26-16.77) between CT1 and CT2, and 3.86% (range: 0.50-13.73) between CT2 and CT3. Table 2 shows the median volume in cc of the 160 Gy isodose for the three scans and the 16 ewes.

Mesh coordinates

Table 3 and Figure 7 shows the center of gravity coordinates of the meshes between CT1 and CT3. Twelve out of 16 animals had a coordinate deviation ≤ 10 mm in the three axes, 1 had one axis > 10 mm, and 3 had 2 axes > 10 mm. The absolute median offsets were 6.3 mm (range: 0.2-13.4) in X, 5.3 mm (range: 0.0-16.8) in Y, and 5.8 mm (range: 0.3-21.7) in Z.

Discussion

Intra-operative external beam radiotherapy and perioperative BT allows dose escalation, while avoiding excessive doses on surrounding normal sensitive tissues [1,2,3,4]. Dose escalation with external beam radiotherapy may be difficult to establish because the high-risk residual volume cannot always be precisely defined on imaging, and/or organs at risk (OARs) cannot be correctly protected from high doses, even with intensity modulated radiation therapy. This is even more challenging in the event of recurrence after previous radiotherapy [2,3,4]. Perioperative radiation therapy with an external beam or BT can overcome these limitations, allowing the dose to be targeted to a limited volume of tissue on the high-risk region defined by a surgeon and radiation oncologist at the time of surgery. However, IORT may be difficult to apply because the facilities are lacking or because the existing applicators are not suited for treating irregular or curved surfaces.
In this setting, BT is a useful option. Temporary iridium 192 implants have been widely used in the extremities or in retro-peritoneal sarcomas [8,9,10,11,12]. The technique is much more complex for pelvic locations than for tumors of the extremities owing to the complex anatomy with curved surfaces and adjacent critical structures. This limitation can be overcome by using HDR intra-operative BT, the dose being delivered perioperatively. Thus, any critical structures may be moved away from the treatment area or protected by shielding [13]. However, there are numerous drawbacks. First, the delivery of a single high-dose can increase the risk of side effects compared to fractionated treatment, especially if any neurologic structures are within the target volume. Second, the operative time can increase significantly, if large volumes are concerned. Third, this type of brachytherapy must be performed in a dedicated shielded room.
I-125 permanent implants can overcome the limitations or drawbacks of the other BT techniques, and be of particular value in the treatment of pelvic locally advanced or recurrent tumors, especially if the recurrence occurs in a previously irradiated region. The advantage of applying this type of BT is the low energy of the I-125; the surrounding normal tissues can be protected more efficiently than with an Ir-192 source, and the much lower dose rate limits the occurrence of side effects. From a technical point of view, the smallness of the seeds, which can be embedded in a flexible mesh, allows any type of irregular or curved anatomical surface to be treated such as the pelvic or sacral wall, with no limitation of the irradiated volume. Moreover, the treatment can begin as soon as the operation has begun, with no need to move the patient to a dedicated unit for radioprotection or to remove catheters. Clinical experience with this type of brachytherapy is scarce and has mostly concerned thoracic tumors and abdominopelvic locations [14,15,16,17,18,19,20].
Since dose delivery with permanent implants lasts 6 months, we needed to verify whether the geometry of the implant could be maintained throughout with no loss of seeds, and that the dose delivered in an experimental study was the same as the prescribed dose, before applying it clinically. We decided to use non-radioactive seeds placed on a 10 cm2 mesh to deliver a theoretical 160 Gy dose at 5 mm, if radioactive I-125 were used. The mesh was then inserted on a pelvic wall of 10 adult ewes by a laparotomy performed under general anesthesia. Three CT scans were performed at day 15, day 70, and day 180 after implantation, corresponding to the post-operative evaluation, to the evaluation at the theoretical half-life of the seeds and finally to the time necessary to deliver 90% of the dose. Of the 10 animals implanted in the first phase of the study, 9 were examined until the last scan. No seed loss was observed in any of the 9 animals. In 8 of 9 animals, the tissue volume receiving 160 Gy was within ±10%, thereby allowing 7 more animals to be included. In the 16 animals investigated at the end of the study, no loss of seeds was observed. There was no movement of the center of gravity of the mesh according to the bony landmarks exceeding 10 mm between the first and the third CT scan for 12 ewes out of 16, the maximum median value being 6.3 mm in the X axis. Moreover, 13 of the 16 animals received the treatment according to the dosimetric constraints, with ≤ 10% differences in the 160 Gy isodose line volume. Thus, the complete resorption of the polyglactin mesh in 8 weeks showed that tissue reworking is sufficient for our study and minimize the seeds migration. Furthermore, TG-43 applies perfectly in homogeneous conditions as our study. This formalism has been adopted by the entire scientific community; however, this method is not based on a model and is closely related to the superposition principle, which limits are known [21,22].
This study has several limitations. First, the abdominal volume of the animals varied owing to the size of their rumen, so it was not possible to obtain dose volume histograms of digestive or other intra-pelvic OAR. Second, it is difficult to position animals for a CT scan even when they are sedated. However, the results show that the technique can be reproduced in mobile animals. The seeds remained in position during the 6 months, with no seed loss and very little migration of the mesh, so the planned dose could be delivered in more than 80% of the cases. The three animals that did not reach the constraints received 79%, 118%, and 118% of the planned 160 Gy volume, in which the pelvic OAR will need to be moved away from the mesh by epiplooplasty or using expanders [23,24,25,26]. Furthermore, appropriate measures will be required to limit radiation exposure to the staff. For radioprotection reasons, a robot will be used to position the mesh in the pelvis, and we are now designing a procedure to allow the seeds to be placed automatically on the mesh. Once this issue is solved, a phase I study will commence for locally advanced or recurrent pelvic tumors in previously irradiated patients using intra-operative permanent I-125 implants as an adjunct to surgery.

Conclusions

This experimental study shows that I-125 seed implantation in pelvic locations is feasible and allows adequate dose delivery during the 6 months needed to deliver the dose with permanent implants. A phase I study can now begin in patients with locally advanced or recurrent pelvic tumors amenable to surgery in combination with surgical resection.

Disclosure

Authors report no conflict of interest.

References

1. Krengli M, Pisani C, Deantonio L et al. Intraoperative radiotherapy in gynaecological and genito-urinary malignancies: focus on endometrial, cervical, renal, bladder, and prostate cancers. Radiat Oncol 2017; 12: 18.
2. Llewelyn M, Taylor A. Re-irradiation of cervical and endometrial cancer. Curr Opin Oncol 2017; 29: 343-350.
3. Backes FJ, Billingsley CC, Martin DD et al. Does intra-operative radiation at the time of pelvic exenteration improve survival for patients with recurrent, previously irradiated cervical, vaginal, or vulvar cancer? Gynecol Oncol 2014; 135: 95-99.
4. Calvo F, Sole C, Lozano M et al. Gynecologic Oncology Intraoperative electron beam radiotherapy and extended surgical resection for gynecological pelvic recurrent malignancies with and without external beam radiation therapy: Long-term outcomes. Gynecol Oncol 2013; 130: 537-544.
5. Nath R, Bice WS, Butler WM et al. AAPM Recommendations on Dose Prescription and Reporting Methods for Permanent Interstitial Brachytherapy for Prostate Cancer: Report of AAPM Task Group No 137, 2009.
6. Nath R, Anderson LL, Luxton G et al. Dosimetry of interstitial brachytherapy sources: Recommendations of the AAPM Radiation Therapy Committee Task Group No 43. Med Phys 1995; 22: 209-234.
7. Nath R, Dewerd LA, Dezarn WA et al. Guidelines by the AAPM and GEC-ESTRO on the use of innovative brachytherapy devices and applications: Report of Task Group No 167. Med Phys 2016; 43: 3178-3205.
8. Pisters BPWT, Harrison LB, Woodruff JM et al. Tissue sarcomas of the extremity and superficial trunk. J Clin Oncol 1994; 12: 1150-1155.
9. Lla C, Delannes M, Minsat M et al. Low-dose intraoperative brachytherapy in soft tissue sarcomas involving neurovascular structure. Radiother Oncol 2006; 78: 10-16.
10. Muhic A, Hovgaard D, Mørk M et al. Local control and survival in patients with soft tissue sarcomas treated with limb sparing surgery in combination with interstitial brachytherapy and external radiation. Radiother Oncol 2008; 88: 382-387.
11. San Miguel I, San Julian M, Cambeiro M et al. Determinants of toxicity, patterns of failure, and outcome among adult patients with soft tissue sarcomas of the extremity and superficial trunk treated with greater than conventional doses of perioperative high-dose-rate brachytherapy and external beam. Int J Radiat Oncol Biol Phys 2011; 81: 529-539.
12. Holloway CL, Delaney TF, Alektiar KM et al. American Brachytherapy Society (ABS) consensus statement for sarcoma brachytherapy. Brachytherapy 2013; 12: 179-190.
13. Lloyd S, Alektiar KM, Nag S et al. Intraoperative high-dose-rate brachytherapy: An American Brachytherapy Society consensus report. Brachytherapy 2017; 16: 446-465.
14. D’Amato T, Galloway M, Szydlowski G et al. Intraoperative brachytherapy following thoracoscopic wedge resection of stage I lung cancer. Chest 1998; 114: 1112-1115.
15. Voynov G, Heron DE, Lin CJ et al. Intraoperative 125 I Vicryl mesh brachytherapy after sublobar resection for high-risk stage I non-small cell lung cancer. Brachytherapy 2005; 4: 278-285.
16. Chen A, Galloway M, Landreneau R et al. Intraoperative I125 brachytherapy for high-risk stage I non-small cell lung carcinoma. Int J Radiat Oncol Biol Phys 1999; 44: 1057-1063.
17. Fernando HC, Santos RS, Benfield JR et al. Lobar and sublobar resection with and without brachytherapy for small stage IA non-small cell lung cancer. J Thorac Cardiovasc Surg 2005; 129: 261-267.
18. Santos R, Colonias A, Parda D et al. Comparison between sublobar resection and 125 Iodine brachytherapy after sublobar resection in high-risk patients with stage I non-small-cell lung cancer. Surgery 2003; 134: 691-697.
19. Mutyala S, Stewart A, Khan AJ et al. Permanent iodine-125 interstitial planar seed brachytherapy for close or positive margins for thoracic malignancies. Int J Radiat Oncol Biol Phys 2010; 76: 1114-1120.
20. Fairweather M, Wang J, Devlin PM et al. Safety and efficacy of radiation dose delivered via iodine-125 brachytherapy mesh implantation for deep cavity sarcomas. Ann Surg Oncol 2014; 22: 1455-1463.
21. Ferré M, Mailleux H, Pierrat N, Dejean C. Brachytherapy dose calculation. Cancer Radiother 2013; 17: 89-92.
22. Beaulieu L, Tedgren A, Carrier JF et al. Report of the Task Group 186 on model-based dose calculation methods in brachytherapy beyond the TG-43 formalism: Current status and recommendations for clinical implementation. Med Phys 2012; 39: 6208-6236.
23. Leblanc E, Narducci F, Bresson L et al. A new laparoscopic method of bowel radio-protection before pelvic chemoradiation of locally advanced cervix cancers. Surg Endosc 2014; 28: 2713-2718.
24. Choi HJ, Lee HS. Effect of omental pedicle hammock in protection against radiation-induced enteropathy in patients with rectal cancer. Dis Colon Rectum 1995; 38: 276-280.
25. Lechner P, Cesnik H. Abdominopelvic omentopexy: preparatory procedure for radiotherapy in rectal cancer. Dis Colon Rectum 1992; 35: 1157-1160.
26. Uehara S, Oue T, Adachi K et al. Tissue expander placement to prevent the adverse intestinal effects of radiotherapy in malignant pelvic tumors. J Pediatr Hematol Oncol 2016; 38: 78-81.
Copyright: © 2018 Termedia Sp. z o. o. This is an Open Access article distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 International (CC BY-NC-SA 4.0) License (http://creativecommons.org/licenses/by-nc-sa/4.0/), allowing third parties to copy and redistribute the material in any medium or format and to remix, transform, and build upon the material, provided the original work is properly cited and states its license.
 
Quick links
© 2024 Termedia Sp. z o.o.
Developed by Bentus.