Monte Carlo dosimetry for a new 32P brachytherapy source using FLUKA code

Razieh Rajabi, Payvand Taherparvar

J Contemp Brachytherapy 2019; 11, 1: 76–90
Online publish date: 2019/02/18
View full text
Get citation
ENW
EndNote
BIB
JabRef, Mendeley
RIS
Papers, Reference Manager, RefWorks, Zotero
AMA
APA
Chicago
Harvard
MLA
Vancouver
 
Purpose
Dosimetric characterization of a new 32P brachytherapy source was studied and the validity of the FLUKA code to reproduce the dosimetric parameters in a water phantom was evaluated. In addition, dose rate distributions around the 32P source sheathed by a catheter and unsheathed source were investigated in different tissue phantoms.

Material and methods
The new 32P source was modeled using FLUKA Monte Carlo code. According to the AAPM TG-60 recommendations, reference of absorbed dose rate, radial dose function, anisotropy function, and an away-along table for quality assurance purposes inside water phantom were calculated. Moreover, the results of the radial dose function and dose rate were obtained for the sheathed source and unsheathed sources at radial distances in different tissue phantoms: liver, fat tissue, 9-component soft tissue, and 4-component soft tissue.

Results
The calculated dosimetric parameters of the new 32P source by FLUKA code in water phantom agreed well with that of the GEANT4 calculation. The 2D away-along dose results were similar to the GEANT4 simulation for distances less than 0.25 cm, small differences were apparent at long distances from the source. Dose rate evaluation for the sheathed source shows that the presence of a catheter increases the dose values up to 2.11% in comparison with the unsheathed source in water phantom. Our results show that the radial dose function calculated in water, as generalized by AAPM TG-60, differed in tissue, especially at large distances from the source.

Conclusions
This work fully characterizes dosimetric parameters of the sheathed and unsheathed new 32P brachytherapy sources in water and different tissue phantoms by using FLUKA code. The results demonstrate that the dose distribution in water differed from the calculated ones in tissue phantoms due to the densities and atomic composition for tissues that are not taken account by the TG-60 formalism.

keywords:

brachytherapy, AAPM TG-60, dosimetry, 32P, FLUKA

references:
Schaart DR. Sealed catheter-based beta sources for intravascular brachytherapy: novel designs and dosimetric characterization. Delft University Press 2002; 1-11.
Badry H, Oufni L, Ouabi H et al. A Monte Carlo investigation of the dose distribution for 60Co High Dose Rate brachytherapy source in water and in different media. Appl Radiat Isot 2018; 136: 104-110.
Pawlicki T, Scanderbeg DJ, Starkschall G. Hendee’s radiation therapy physics. Wiley-Blackwell 2016; 215-230.
Thomadsen B, Rivard M, Butler W. Brachytherapy physics. Med Phys 2005: 1-5.
Junxiang W, Shihu Y, Jing H et al. Monte Carlo dosimetry of a new 90Y brachytherapy source. J Contemp Brachytherapy 2015; 7: 397-406.
Choi CH, Han HS, Son KJ et al. Dosimetry of a new P-32 ophthalmic applicator. Med Phys 2011; 38: 6143-6151.
Sun L, Zhu X, Xu L et al. Antitumor effects of 32P‑chromic‑poly (L‑lactide) brachytherapy in nude mice with human prostate cancer. Oncol Lett 2013; 6: 687-692.
Lyra ME, Andreou M, Georgantzoglou A et al. Radionuclides used in nuclear medicine therapy – from production to dosimetry. Curr Med Imaging Rev 2013; 9: 51-75.
Junxiang W, Jing H, Fengxiang L et al. Monte Carlo dosimetric parameter study of a new 32P brachytherapy source. Br J Radiol 2016; 89: 1-9.
Anjomrouz M, Bakht MK, Sadeghi M. Feasibility study of FLUKA Monte Carlo simulation for a beta emitting brachytherapy source: dosimetric parameters of 142Pr glass seed. J Radioanal Nucl Chem 2016; 3: 947-953.
Sadeghi M, Taghdiri F, Saidi P. Dosimetric characteristics of the 192Ir high-dose-rate afterloading brachytherapy source. Japan J Radiol 2011; 29: 324-329.
Paiva E. Numerical calculation of relative dose rates from cylindrical 32P and 90Y beta sources used in intravascular brachytherapy. Scientia Plena 2014; 9: 1-12.
Taherparvar P, Sadremomtaz A. Development of GATE Monte Carlo simulation for a CsI pixelated gamma camera dedicated to high resolution animal SPECT. Australas Phys Eng Sci Med 2018; 41: 31-39.
Almansa López JF, Torres Donaire J, Guerrero Alcalde R. Monte Carlo dosimetry of the most commonly used 192Ir high dose rate brachytherapy sources. Rev Fis Med 2011; 12: 159-168.
Kiavar O, Sadeghi M. Modeling and dose calculations of a pure beta emitting 32P coated stent for intracoronary brachytherapy by Monte Carlo code. Iran J Radiat Res 2012; 9: 257-263.
Carter AJ, Laird JR. Experimental results with endovascular irradiation via a radioactive stent. Int J Radiat Oncol Biol Phys 1996; 36: 797-803.
Holmes SM, DeWerd LA, Micka JA. Experimental determination of the radial dose function of 90Sr/90Y IVBT sources. Med Phys 2006; 33: 3225-3233.
Mourtada FA, Soares CG, Seltzer SM et al. Dosimetry characterization of 32P catheter-based vascular brachytherapy source wire. Med Phys 2000; 27: 1770-1776.
Torres J, Buades MJ, Almansa JF et al. Dosimetry characterization of 32P intravascular brachytherapy source wires using Monte Carlo codes PENELOPE and GEANT4. Med Phys 2004; 31: 296-304.
Asenjo J, Fernández-Varea JM, Sánchez-Reyes A. Characterization of a high-dose-rate 90Sr–90Y source for intravascular brachytherapy by using the Monte Carlo code PENELOPE. Phys Med Biol 2002; 47: 697-711.
Wang R, Li XA. A Monte Carlo calculation of dosimetric parameters of 90Sr / 90Y and 192 Ir SS sources for intravascular brachytherapy. Med Phys 2000; 27: 2528-2535.
Cerutti F, Engel R, Fedynitch A et al. New developments in FLUKA. NEA/NSC/R 2015; 3: 220-230.
Cao Z, Ruan XC, Meng BD et al. Determination of dosimetric characteristics for 125I seed source with FLUKA code. Nucl Sci Tech 2014; 25: 1-3.
Demir N, Tarim UA, Popovici MA et al. Investigation of mass attenuation coefficients of water, concrete and bakelite at different energies using the FLUKA Monte Carlo code. J Radioanal Nucl Chem 2013; 2: 1303-1307.
Andersen V, Ballarini F, Battistoni G et al. The application of FLUKA to dosimetry and radiation therapy. Radiat Prot Dosimetry 2005; 116: 113-117.
ICRU Report No. 56. Dosimetry of external beta rays for radiation protection, International Commission on Radiation Units and Measurements, Bethesda, MD, 1997.
Ferrari A, Sala PR, Fasso A et al. Fluka: a multi-particle transport code. CERN 2005.
DeWerd LA, Ibbott GS, Meigooni AS et al. A dosimetric uncertainty analysis for photon-emitting brachytherapy sources: Report of AAPM Task Group No. 138 and GEC-ESTRO. Med Phys 2011; 38: 782- 801.
Chiu-Tsao ST, Schaart DR, Soares CG et al. Dose calculation formalisms and consensus dosimetry parameters for intravascular brachytherapy dosimetry: Recommendations of the AAPM Therapy Physics Committee Task Group No. 149. Med Phys 2007; 34: 4126-4157.
Xu Z, Almond PR, Deasy JO. The dose distribution produced by a 32P source for endovascular irradiation. Int J Radiat Oncol Biol Phys 1996; 36: 933-939.
Jursinic PA. Quality assurance measurements for high-dose-rate brachytherapy without film. J Appl Clin Med Phys 2014; 15: 246-261.
Aalbers AHL, De Brabandere M, Koedooder C et al. Dosimetry and quality control of brachytherapy with low-energy photon sources (125I). The Netherlands Commission on Radiation Dosimetry Subcommittee 2012.
Nath R, Amols H, Coffey C et al. Intravascular brachytherapy physics: Report of the AAPM Radiation Therapy Committee Task Group No. 60. Med Phys 1999; 26: 119-152.
Ghorbani M, Salahshour F, Haghparast A et al. Effect of tissue composition on dose distribution in brachytherapy with various photon emitting sources. J Contemp Brachytherapy 2014; 6: 54-67.
ICRU. ICRU Report No. 44, Tissue substitutes in radiation dosimetry and measurement. ICRU, Bethesda 1989.
 
Quick links
© 2019 Termedia Sp. z o.o. All rights reserved.
Developed by Bentus.
PayU - płatności internetowe