Original paper
Mapping of dose distribution from IMRT onto MRI-guided high dose rate brachytherapy using deformable image registration for cervical cancer treatments: preliminary study with commercially available software

Hayeon Kim
,
M. Saiful Huq
,
Chris Houser
,
Sushil Beriwal
,
Dariusz Michalski

J Contemp Brachytherapy 2014; 6, 2: 178–184
Online publish date: 2014/06/03
Article file
- mapping of dose.pdf  [0.22 MB]
Get citation
 
 

Purpose

The standard treatment for locally advanced cervical cancer is to deliver a combination of external beam radiotherapy (EBRT) and brachytherapy as a boost to primary tumor and cervix [1]. High dose rate (HDR) brachytherapy combined with image guided brachytherapy (IGBT) has recently been introduced to treat locally advanced cervical cancer and is now gaining popularity [2,3]. This emerging technique has been shown to provide improvements in dose volume parameters and in clinical outcome [4-8]. In 2005, GEC-ESTRO working group published recommendations for IGBT for cervical cancer. These recommendations are based primarily on the use of repetitive magnetic resonance imaging (MRI) performed at the time of the delivery of each fraction of brachytherapy [9-11]. Magnetic resonance imaging- based IGBT facilitates better delineation of target volume. Dose optimization can then be performed such that the classical pear shaped isodose lines better conform to the shape of the well-delineated target volume. This improvement in target dose coverage and reduction in dose to critical organs have shown to improve local control and decreased toxicities [12,13]. To provide consistency in dose reporting, the GEC-ESTRO working group recommended that for MRI-based IGBT of cervical cancer, the total biological dose should be calculated by linearly adding the equivalent dose in 2 Gy fractions (EQD2) resulting from EBRT and HDR brachytherapy [9]. The ongoing international study on MRI-guided brachytherapy for locally advanced cervical cancer, EMBRACE, has adopted these recommendations for dose reporting [14,15].
The current practice for the calculation of total biological dose is to assume that the dose distribution from EBRT is homogenous. For example, all patients who received 45 Gy EBRT would be considered to have a uniform EQD2 (43.2 Gy for organs at risk and 44.3 Gy for tumor) for calculation of the total equivalent dose, regardless of the heterogeneity of dose that is presumed to exist in the EBRT plan. To calculate the total equivalent dose from combined modality treatments, linear addition of EQD2 from each modality is often used [9]. However, this simple linear addition of the dose matrices is based on the rigid image registration. It does not align deforming anatomy adequately [16]. A more accurate methodology would be to perform a deformable registration between the two imaging modalities, CT, and MRI, which will enable a mapping of each dose voxel from one modality to another by taking deformed anatomy into account [16,17].
In this study, we utilized commercially available deformable image registration software to calculate the deformed IMRT dose voxel by voxel, mapped on the MRI image for every brachytherapy fraction to estimate the EBRT dose distribution. Additionally, the deformed IMRT dose mapped on MR images is compared to the current clinical practice (uniform dose distribution from EBRT). To the best of our knowledge, such a study focusing on cervical cancer treatment has not been reported in the literature.

Material and methods

Patient data and treatment planning

Nine cervical cancer patients, FIGO stage IB1 to IIIB, were retrospectively analyzed. All patients were treated to a total dose of 45 Gy in 25 fractions using IMRT, followed by MRI-guided 3D HDR brachytherapy (BT) using a ring and tandem applicator. For three patients, the pelvic and/or para-aortic nodes were treated to a total dose of 55 Gy in 25 fractions using IMRT with simultaneous integral boost (SIB) technique.
Patients underwent a CT simulation for treatment planning in supine position with full bladder and empty rectum without a vaginal marker. For all patients, two clinical target volumes (CTV) were created. The CTV1 included common iliac, external iliac, and internal iliac lymph nodes regions in all patients. The inguinofemoral and para-aortic nodes were included only as clinically indicated. The CTV2 included the uterus, cervix, upper vagina, and parametria. The CTV1 was expanded by 0.7 cm and CTV2 was expanded by 1-2 cm, and combined together to create the final PTV. Brachytherapy was initiated during the fourth or fifth week of external beam treatment, following the placement of a Smit sleeve. Brachytherapy was carried out once a week until the completion of EBRT, at which point it was delivered twice a week. The CT/MRI compatible ring and tandem applicator (Nucletron, an Elekta company, Elekta AB, Stockholm, Sweden) with a tandem length of 40 or 60 mm, curvature of 45o or 60o, and a ring diameter of 26, 30, or 34 mm was used for brachytherapy. A rectal retractor was used, and a Foley catheter was inserted in all patients. This ensures consistent and reproducible setup for all patients for each fraction. Orthogonal X-ray films were taken for the verification of applicator position reproducibility for each fraction prior to MRI. All nine patients underwent a repetitive MRI (GE SIGNA HD platform 1.5 T) for planning prior to each fraction of brachytherapy. With the applicator in place, T2-weighted sequences in the axial, coronal, and sagittal planes were acquired using 2.0 mm slices. Magnetic resonance imaging-based 3D treatment plans were generated for each of the five fractions for all patients in the study (total of 45 fractions). For each fraction, the high risk clinical target volume (HR-CTV) and organs at risk (OARs) were delineated following the GEC-ESTRO recommendations, and then transferred to Nucletron Plato Brachytherapy Planning System™ Version 14.3 (Nucletron, an Elekta company, Elekta AB, Stockholm, Sweden). Each patient was evaluated for the gross tumor volume (GTV), which is a part of HR-CTV, with MR images before treatment course and after treatment of external beam. At the time of brachytherapy, physicians evaluated the GTV and HR-CTV based on GEC-ESTRO contouring recommendations. Point A (traditional dose prescription point defined as 2.0 cm above and lateral to the cervical os) was marked on corresponding slices on MR images as a reference starting point for dose optimization. After standard loading of source dwell positions and weighting in tandem and ring, manual optimization was performed based on isodose lines in axial slices of MR images, and evaluation of dose-volume histogram (DVH) constraints for HR-CTV and OARs including rectum, bladder, and sigmoid. The total dose of 27.5-30.0 Gy delivered in 5 fractions (5.5-6.0 Gy/fraction) was prescribed to the HR-CTV. Plans were designed to achieve an EQD2 of 80-85 Gy for HR-CTV and limit dose to 2 cc (D2cc) of bladder, rectum, and sigmoid to be ≤ 85 Gy, ≤ 70 Gy and ≤ 70 Gy, respectively.

Conventional EBRT EQD2 dose calculation – uniform dose distribution method

All prescribed and calculated physics doses to the target and OARs were converted to EQD2 using the linear quadratic model with the assumption of α/β = 3 for OAR and α/β = 10 for HR-CTV. As per GEC-ESTRO recommendations, the dose distribution of EBRT to the HR-CTV and OARs was assumed to be uniform. Thus, EQD2 to HR-CTV and OARs was always reported as 44.3 Gy and 43.2 Gy, respectively.

Deformable registration and non-uniform EBRT dose distribution method

To map the actual dose distribution on EBRT CT images onto the OARs and HR-CTV contoured on the MRI for brachytherapy, image registration was required. The EBRT planning CT images were fused retrospectively to the MR images acquired for each fraction of brachytherapy using commercially available software (Velocity AI 2.7., Velocity Medical Solutions, Atlanta, GA, USA). First, a rigid image registration was performed using the bony anatomy and intact uterus as landmarks. Then, EBRT planning CT images were deformed to match to each MR image (five MR image set per patient) using multi-pass iterative optimization using B-Spline deformable image registration algorithm. This algorithm aligns images by maximizing the mutual information. The deformation vector field (DVF) created for the image registration allows Velocity to deform dose distribution voxel by voxel from IMRT on MR images. For each fraction, the deformed dose distribution from IMRT on MR image was then calculated by warping doses based on the quality of the DVF obtained during image registration [18]. We used a qualitative evaluation process to ensure that the deformed images were reasonable. That is, these fused images were visually inspected to identify physically unrealistic deformations such as folds and tears. Also, the displacement (mapping) vectors were reviewed to check the direction and magnitude of the displacement from the deformed images to the reference images [19]. In addition, warping volume histogram was evaluated to estimate the amount of warping voxels across the deformed OARs. The deformed EBRT dose voxel by voxel was calculated for HR-CTV, rectum, bladder, and sigmoid for each fraction of MRI brachytherapy. Dose-volume histogram for each organ was generated for each fraction from the deformed EBRT dose and converted to EQD2.

Statistical analysis

Paired sample t-tests were conducted for the comparison between uniform dose assumption and deformed dose. Additionally, correlated data analysis for repeated measures over 5 fractions within a subject was used to identify random effects. Statistical analyses were performed with STATA software version 12.1 (STATA Corp, College Station, TX, USA), and two-tailed p-values of < 0.05 were considered to be significant.

Results

The mean HR-CTV ± SD was 35.7 ± 13.2 cc (range: 18-53 cc). The mean EQD2 ± SD of the deformed EBRT dose for HR-CTV D90 (dose covering 90% of the volume) was 45.7 ± 0.7 Gy (range: 45.2-47.0) and was significantly greater than that obtained using a uniform EBRT dose distribution (44.3 Gy, p < 0.001). The D2cc for bladder, rectum, and sigmoid was 46.4 ± 1.2 Gy (range: 45.0-49.0), 46.2 ± 1.0 Gy (range: 44.5-48.0), and 48.0 ± 2.5 Gy (range: 46.0-52.8), respectively based on deformed EBRT doses. These doses were significantly higher than the D2cc calculated based on a uniform dose distribution (43.2 Gy for all OARs, p < 0.001). For all patients, the doses obtained using deformable image registration method was found to be consistently higher than those obtained using uniform dose distribution assumption. We then sought to explore whether the doses to OAR and HR-CTV differed with and without the use of SIB technique with IMRT. The comparison of doses for bladder, rectum, and sigmoid treated with and without SIB were 46.9 Gy vs. 45.4 Gy, 46.4 Gy vs. 45.4 Gy, and 49.3 Gy vs. 46.0 Gy, respectively. The HR-CTV dose difference was not found to be as pronounced (46.0 Gy for SIB vs. 45.2 Gy without SIB).
Additionally, the total EQD2 from the entire course of therapy were recalculated and compared with the current practice. The mean total EQD2 of the HR-CTV between deformed vs. uniform dose was 84.7 Gy vs. 83.5 Gy (p = 0.43). The mean total EQD2 to 2cc of the bladder for deformed and uniform dose was 77.2 Gy vs. 74.4 Gy (p = 0.31). Similarly, the mean total EQD2 to 2cc of the rectum and sigmoid for deformed and uniform dose was 58.7 Gy vs. 56 Gy (p = 0.12) and 68.7 Gy vs. 64.7 Gy (p = 0.18), respectively. Table 1 and Table 2 summarize the mean deformed EBRT EQD2 and total EQD2 for all patients, and doses for each individual, respectively.
Interestingly, one patient (11%) had > 85 Gy for total EQD2 to D2cc of the bladder, and three patients (33%) had > 70 Gy for sigmoid when the deformed dose applied. However, based on conventional dose calculation with uniform EBRT dose distribution, none were found to receive higher than 85 Gy and 70 Gy for the total EQD2 to D2cc of the bladder and sigmoid.
For each patient, repeated measures of dose for all OARs over 5 fractions were found to be highly correlated within a subject (bladder, rectum, and sigmoid; p < 0.001). Therefore, random effects between fractions were not found within a patient.

Discussion

The present study demonstrates that in cervical cancer patients treated with a combination of external beam radiation therapy (IMRT in the present study) and MRI-based image guided HDR brachytherapy, the contribution of HR-CTV and OAR doses from IMRT to total dose vary between patients with a deformable image registration method. Our findings suggest that unlike the current GEC-ESTRO recommendations, where the EBRT dose is assumed be uniform (with an EQD2 of 44.3 Gy for HR-CTV and 43.2 Gy for the OARs for a 45 Gy EBRT prescription) for all patients, we found that deformable registration revealed significant differences. As shown in Table 3, the heterogeneous IMRT dose distributions across PTV area exist. Importantly, we discovered that for 33% of patients in the study (n = 3), the deformable dose calculation method identified that OAR doses exceeded GEC-ESTRO recommendations, but conventional calculation assuming EBRT dose uniformity did not identify this. Much larger further studies are needed to confirm these differences and determine the clinical significance of this.
However, Van de Kamer et al. [17] has shown that a homogeneous EBRT dose is a still good approximation for combining the dose from EBRT and brachytherapy. It should be noted that in this study, two brachytherapy fractions that combined intra-cavitary and interstitial pulsed dose rate (PDR) brachytherapy after the EBRT were delivered, and only one fraction of MR image from brachytherapy was fused with the CT images from EBRT by using 3D biological modeling based on the registration with bony anatomy matching. This is different from our clinical practice that consists of delivering five fractions of brachytherapy using only an intra-cavitary applicator after EBRT with subsequent deformable image registration applied to all 5 fractions of MR images from brachytherapy.
Although the mean differences between a homogeneous dose assumption and deformed EBRT dose contribution is statistically significant in the present study, the mean total EQD2 between homogeneous EBRT dose and deformed dose was not significant, and was less than 10% for all OARs and target. This results in concordance with those from the above study. Thus, while there were no significant differences between the two dose calculation methods when considering the entire treatment course, for a proportion of patients, deformable dose mapping may have altered the brachytherapy plan.
Andersen et al. [20] and Jamema et al. [21] reported that D2cc of bladder and rectum in MRI-based image guided brachytherapy of cervical cancer is a stable spatial location for direct dose summation for each fraction. Our total D2cc for OARs from the contribution of each brachytherapy fraction is based on the same observation as their study.
The present study also shows that for those patients who were treated using SIB technique, the OAR doses from IMRT were higher than the corresponding doses treated without using the SIB technique. Sigmoid dose had higher difference than ones from bladder and rectum because of proximity of these organs to the pelvic lymph nodes. None of patients in the present study received the total EQD2 to 2cc of the bladder dose > 85 Gy with homogeneous dose assumption, but one patient had 86 Gy after deformed IMRT dose was applied. Similarly for sigmoid dose, three patients had > 70 Gy after the deformed IMRT dose was applied, while none of patients received > 70 Gy with a homogeneous dose distribution. These patients were treated using SIB technique. Since the D2cc of bladder, rectum, and sigmoid is a good predictive value for the normal tissue toxicity [22], it is important to quantify these DVH parameters accurately.
Our preliminary results have several limitations. First, the evaluation of deformable image registration accuracy was based on the qualitative methods, not quantitative assessment. There is a lack of straightforward metrics to evaluate deformable image registration errors [23], although deformable image registration algorithms are widely accepted and implemented in clinical practice [23-28]. Thus, visual inspection (qualitative evaluation) is still the most widely used evaluation method [19]. In those studies, the evaluation of deformable image registration did not deal with the combined modality of external beam and the complex brachytherapy. Secondly, we excluded patients who had unrealistic warping maps which were associated with unreasonable deformation. We initially started retrospective review with 15 patients for deformable image registration, but 9 out of them were reasonable based on our assessment. Due to large organ deformations between external beam treatment and brachytherapy fraction with applicator in situ, image registration causes extreme deformed organs and lead to incorrect dose mapping. Although there are commercially available software packages which have been widely accepted in clinical use, the degree of deformation associated with cervical cancer makes this site particularly challenging [25]. Thirdly, sigmoid D2cc assessment is highly uncertain due to mobility of this organ. In IMRT planning CT, most part of sigmoid is close to PTV area. Thus we assumed that sigmoid D2cc (representing hot spot/high dose area) from IMRT plan is close to PTV area, and the D2cc for sigmoid at the time of brachytherapy would be coincided with the PTV area based on the maximized mutual information using deformed image registration. Additionally, we did not investigate the impact of using deformable image registration between brachytherapy fractions. Brachytherapy linear dose addition between fractions without deformable image registration is based on the assumption that the hot spots are located in the same region of each organ for each fraction that results in an overestimation of dose [29]. Thus, the maximum dose to OARs determined using deformable image registration cannot be higher than the linear dose addition method, because the hot spots do not necessarily overlap for each fraction. Finally, we did not include the effects of inter fraction organ motion during the course of EBRT and used static CT scans at the time of planning for deformation registration and dose calculations. Since daily setup uncertainty causes the variation of the dose distribution from the planning, these factors also need to be accounted for in the deformed dose from EBRT [30].

Conclusions

The present study assessed the composite dose distribution of EBRT and brachytherapy in cervical cancer treatments with a commercially available deformable image registration method. This study demonstrated that cervical cancer patients treated with a combination of external beam radiation therapy (IMRT in the present study) and MRI-based image guided HDR brachytherapy the contribution of HR-CTV and OAR doses from IMRT to total dose vary from one patient to another, and it is prominent in cases where SIB boost is delivered to pelvic nodes. Therefore, the assumption that EBRT contributes the same dose value of HR-CTV and OARs for all patients (an EQD2 of 44.3 Gy for HR-CTV and 43.2 Gy for the OARs) may need to be carefully investigated further based on deformable image registration.

Acknowledgements

We would like to thank Sharon Callaway from Velocity Medical Solutions for advice regarding the VelocityAI software.

Disclosure

Authors report no conflict of interest.

References

1. Nag S, Erickson B, Thomadsen B et al. The American Brachytherapy Society recommendations for high-dose-rate brachytherapy for carcinoma of the cervix. Int J Radiat Oncol Biol Phys 2000; 48: 201-211.
2. Dimopoulos JC, De Vos V, Berger D et al. Inter-observer comparison of target delineation in MRI based cervix cancer brachytherapy according to the GYN GEC-ESTRO recommendations. Radiother Oncol 2007; 83: 18-19.
3. Viswanathan AN, Erickson BA. Three-dimensional imaging in gynecologic brachytherapy: a survey of the American Brachytherapy Society. Int J Radiat Oncol Biol Phys 2010; 76: 104-109.
4. Pötter R, Dimopoulos J, Georg P et al. Clinical impact of MRI assisted dose volume adaptation and dose escalation in brachytherapy of locally advanced cervix cancer. Radiother Oncol 2007; 83: 148-155.
5. Dimopoulos JC, Pötter R, Lang S et al. Dose-effect relationship for local control of cervical cancer by magnetic resonance image-guided brachytherapy. Radiother Oncol 2009; 93: 311-315.
6. Wang B, Kwon A, Zhu Y et al. Image-guided intracavitary high-dose-rate brachytherapy for cervix cancer: a single institutional experience with three-dimensional CT-based planning. Brachytherapy 2009; 8: 240-247.
7. Tan LT, Coles CE, Hart C, et al. Clinical impact of computed tomography-based image-guided brachytherapy for cervix cancer using the tandem-ring applicator – the Addenbrooke’s experience. Clin Oncol (R Coll Radiol) 2009; 21: 175-182.
8. Beriwal S, Kannan N, Kim H et al. Three dimensional high dose rate intracavitary image guided brachytherapy for the treatment of cervical cancer using a hybrid magnetic resonance imaging/computed tomography approach: feasibility and early results. Clin Oncol 2011; 23: 685-690.
9. Pötter R, Haie-Meder C, Van Limbergen E et al. Recommendations from Gynaecological (GYN) GEC ESTRO Working Group (II): Concepts and terms in 3D image-based treatment planning in cervix cancer brachytherapy-3D dose volume parameters and aspects of 3D image-based anatomy, radiation physics, radiobiology. Radiother Oncol 2006; 78: 67-77.
10. Lang S, Nulens A, Briot E et al. Intercomparison of treatment concepts for MR image assisted brachytherapy of cervical carcinoma based on GYN GEC-ESTRO recommendations. Radiother Oncol 2006; 78: 185-193.
11. Haie-Meder C, Pötter R, Van Limbergen E et al. Recommendations from Gynaecological (GYN) GEC-ESTRO Working Group (I): Concepts and terms in 3D image based 3D treatment planning in cervix cancer brachytherapy with emphasis on MRI assessment of GTV and CTV. Radiother Oncol 2005; 74: 235-245.
12. Georg P, Lang S, Dimopoulos JC et al. Dose-volume histogram parameters and late side effects in magnetic resonance image-guided adaptive cervical cancer brachytherapy. Int J Radiat Oncol Biol Phys 2011; 79: 356-362.
13. Tharavichitkul E, Wanwilairat S, Chakrabandhu S et al. Image- guided brachytherapy (IGBT) combined with whole pelvic intensity-modulated radiotherapy (WP-IMRT) for locally advanced cervical cancer: a prospective study from Chiang Mai University Hospital, Thailand. J Contemp Brachytherapy 2013; 5: 10-16.
14. Pötter R, Kirisits C, Fidarova EF et al. Present status and future of high-precision image guided adaptive brachytherapy for cervix carcinoma. Acta Oncol 2008; 47: 1325-1336.
15. Pötter R, Fidarova E, Kirisits C et al. Image-guided adaptive brachytherapy for cervical carcinoma. Clin Oncol 2008; 20: 426-432.
16. Vásquez Osorio EM, Hoogeman MS, Teguh DN et al. Three dimensional dose addition of external beam radiotherapy and brachytherapy for oropharyngeal patients using nonrigid registration. Int J Radiat Oncol Biol Phys 2011; 80: 1268-1277.
17. Van de Kamer JB, De Leeuw AA, Moerland MA et al. Determining DVH parameters for combined external beam and brachytherapy treatment: 3D biological dose adding for patients with cervical cancer. Radiother Oncol 2010; 94: 248-253.
18. Lin H, Ayan A, Zhu T et al. A quantitative evaluation of velocity AI deformable image registration. Med Phys 2010; 37: 3126.
19. Gao S, Zhang L, Wang H et al. A deformable image registration method to handle distended rectums in prostate cancer radiotherapy. Med Phys 2006; 33: 3304-3312.
20. Andersen ES, Noe KØ, Sørensen TS et al. Simple DVH parameter addition as compared to deformable registration for bladder dose accumulation in cervix cancer brachytherapy. Radiother Oncol 2013; 107: 52-57.
21. Jamema SV, Mahantshetty U, Tanderup K et al. Inter-application variation of dose and spatial location of D2cc volumes of OARs during MR image based cervix brachytherapy. Radiother Oncol 2013; 107: 58-62.
22. Georg P, Kirisits C, Goldner G et al. Correlation of dose-volume parameters, endoscopic and clinical rectal side effects in cervical cancer patients treated with definitive radiotherapy including MRI-based brachytherapy. Radiother Oncol 2009; 91: 173-180.
23. Schreibmann E, Pantalone P, Waller A et al. A measure to evaluate deformable image registration fields in clinical settings. J Appl Clin Med Phys 2012; 13: 126-139.
24. Lawson JD, Schreibmann E, Jani AB et al. Quantitative evaluation of a cone beam computed tomography-planning computed tomography deformable image registration method for adaptive radiation therapy. J Appl Clin Med Phys 2007; 8: 96-113.
25. Bondar L, Hoogeman MS, Vásquez Osorio EM et al. A symmetric nonrigid registration method to handle large organ deformation in cervical cancer patients. Med Phys 2010; 37: 3760-3772.
26. Stanley N, Zhong H, Glide-Hurst C et al. Systematic evaluation of a deformable image registration algorithm from a commercial software package. Med Phys 2012; 39: 3876.
27. Nie K, Chuang C, Kirby N et al. Site specific deformable image registration algorithm selection using patient-based simulated deformations. Med Phys 2013; 40: 041911.
28. Kirby N, Chuang C, Ueda U et al. The need for application- based adaptation of deformable image registration. Med Phys 2013; 40: 011702.
29. Kirisits C, Pötter R, Lang S et al. Dose and volume parameters for MRI based treatment planning in intracavitary brachytherapy for cervical cancer. Int J Radiat Oncol Biol Phys 2005; 62: 901-911.
30. Tanderup K, Nesvacil N, Pötter R et al. Uncertainties in image guided adaptive cervix cancer brachytherapy: impact on planning and prescription. Radiother Oncol 2013; 107: 1-5.
Copyright: © 2014 Termedia Sp. z o. o. This is an Open Access article distributed under the terms of the Creative Commons Attribution-NonCommercial-ShareAlike 4.0 International (CC BY-NC-SA 4.0) License (http://creativecommons.org/licenses/by-nc-sa/4.0/), allowing third parties to copy and redistribute the material in any medium or format and to remix, transform, and build upon the material, provided the original work is properly cited and states its license.
 